Accuracy of the Welch Allyn Vital Signs Monitor 52000 automatic blood pressure measuring device according to a modified British Hypertension Society protocol

被引:12
作者
Braam, RL [1 ]
de Maat, C [1 ]
Thien, T [1 ]
机构
[1] Univ Med Ctr St Radboud Nijmegen, Dept Internal Med, Div Hypertens & Vasc Pathol, Nijmegen, Netherlands
关键词
blood pressure; device validation; oscillometric monitor; British Hypertension Society Protocol;
D O I
10.1097/00126097-200206000-00007
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
The accuracy of the Welch Allyn Vital Signs Monitor, a compact device for the oscillometric measurement of blood pressure, was determined according to the British Hypertension Society protocol. The monitor achieved a grade C for diastolic and a grade D for systolic blood pressure. The device is suitable for monitoring a patient, for example post-operatively, in the emergency department or during an intervention. The device cannot, however, be recommended for an exact determination of blood pressure when compared with the mercury sphygmomanometer. In an earlier validatation report, the Welch Allyn Vital Signs Monitor achieved a grade A for both diastolic and systolic blood pressure. After adjusting for the difference in method of calculating the grades used in the two studies, there remained a considerable difference in grading results, for which no clear reason could be found.
引用
收藏
页码:185 / 189
页数:5
相关论文
共 12 条
[1]   ASSESSMENT OF OSCILLOMETRIC NONINVASIVE BLOOD-PRESSURE MONITORS USING THE DYNATECH NEVADA CUFFLINK ANALYZER [J].
AMOORE, JN .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1993, 17 (01) :25-31
[2]  
[Anonymous], 1987, AM NAT STAND EL AUT
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   The influence of the method of data analysis on the reported accuracy of automated blood pressure measuring devices [J].
Braam, RL ;
Thien, T .
JOURNAL OF HYPERTENSION, 2001, 19 (10) :1765-1767
[5]   Are aneroid sphygmomanometers accurate in hospital and clinic settings? [J].
Canzanello, VJ ;
Jensen, PL ;
Schwartz, GL .
ARCHIVES OF INTERNAL MEDICINE, 2001, 161 (05) :729-731
[6]   Validation of the Welch Allyn 'Vital Signs' oscillometric blood pressure monitor [J].
Jones, CR ;
Taylor, K ;
Poston, L ;
Shennan, AH .
JOURNAL OF HUMAN HYPERTENSION, 2001, 15 (03) :191-195
[7]   Proposals for simplifying the validation protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation [J].
O'Brien, E .
BLOOD PRESSURE MONITORING, 2000, 5 (01) :43-45
[8]   Will the millimetre of mercury be replaced by the kilopascal? [J].
O'Brien, E .
JOURNAL OF HYPERTENSION, 1998, 16 (03) :259-261
[9]   Replacing the mercury sphygmomanometer [J].
O'Brien, E .
BRITISH MEDICAL JOURNAL, 2000, 320 (7238) :815-816
[10]   State of the market in 2001 for blood pressure measuring devices [J].
O'Brien, E .
BLOOD PRESSURE MONITORING, 2001, 6 (04) :171-176