Clinical comparison of BACTEC 9240 plus aerobic/F resin bottles and the isolator aerobic culture system for detection of bloodstream infections

被引:26
作者
Cockerill, FR
Reed, GS
Hughes, JG
Torgerson, CA
Vetter, EA
Harmsen, WS
Dale, JS
Roberts, GD
Ilstrup, DM
Henry, NK
机构
[1] MAYO CLIN & MAYO FDN,DEPT INTERNAL MED,DIV INFECT DIS,ROCHESTER,MN 55905
[2] MAYO CLIN & MAYO FDN,DEPT LAB MED & PATHOL,DIV CLIN CHEM & IMMUNOL,ROCHESTER,MN 55905
[3] MAYO CLIN & MAYO FDN,DEPT HLTH SCI RES,BIOSTAT SECT,ROCHESTER,MN 55905
[4] MAYO CLIN & MAYO FDN,DEPT PEDIAT & ADOLESCENT MED,ROCHESTER,MN 55905
关键词
D O I
10.1128/JCM.35.6.1469-1472.1997
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
The Plus Aerobic/F resin bottle of the BACTEC 9240 automated blood culture system (Becton Dickinson Diagnostic Instrument Systems, Sparks, Md.) was compared with aerobic culture of the Isolator system (Wampole Laboratories, Cranbury, N.J.) for the detection of bloodstream microorganisms from 6,145 blood cultures collected from adult patients with suspected septicemia. The BACTEC resin bottles were incubated for 7 days, and the sediment from the Isolator tube was inoculated to sheep blood and chocolate agars which were incubated for 72 h and to inhibitory mold, brain heart infusion, and Sabouraud agars which were incubated for 21 days. A total of 622 microorganisms were recovered from 583 blood cultures. The BACTEC resin bottle recovered statistically significantly more pathogens overall than the Isolator system (P = 0.0006). When individual pathogens isolated from either system for a 7-day study period were assessed, it was determined that the BACTEC resin bottle detected statistically significantly more isolates of Staphylococcus aureus (P = 0.0113) and coagulase-negative Staphylococcus spp. (P = 0.0029) than the Isolator system. The BACTEC resin bottle also detected statistically significantly more bloodstream infections (septic episodes) caused by coagulase-negative Staphylococcus spp. (P = 0.0146). The Isolator system recovered statistically significantly more contaminants overall (P < 0.0001), and among this group of microorganisms, recovered statistically significantly more Bacillus spp. (P < 0.0001), coagulase-negative Staphylococcus spp, (P < 0.0001), and viridans group Streptococcus spp, (P = 0.0156). The Isolator system detected statistically significantly more isolates of Histoplasma capsulatum (P = 0.004), but all of these isolates were detected at greater than or equal to 7 days of incubation of fungal plates, i.e., after the system to system comparison study period (7 days). In blood culture sets which produced growth of the same pathogen in both systems, there was a statistically significant difference in median time to detection for all pathogens combined favoring the BACTEC resin bottle over the Isolator tube (P < 0.05). When assessing individual microorganisms, the median times for detection of S. aureus, Enterococcus spp., and Pseudomonas spp. were all statistically significantly less for the BACTEC system (P < 0.05). The BACTEC instrument had 79 (1.3%) false positive signals. The BACTEC system required less processing time than the Isolator system and eliminates the hands-on time for detection of positive cultures required with the Isolator system.
引用
收藏
页码:1469 / 1472
页数:4
相关论文
共 7 条
  • [1] CONTAMINANT BLOOD CULTURES AND RESOURCE UTILIZATION - THE TRUE CONSEQUENCES OF FALSE-POSITIVE RESULTS
    BATES, DW
    GOLDMAN, L
    LEE, TH
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1991, 265 (03): : 365 - 369
  • [2] Clinical comparison of Difco ESP, Wampole Isolator, and Becton Dickinson Septi-Chek aerobic blood culturing systems
    Cockerill, FR
    Torgerson, CA
    Reed, GS
    Vetter, EA
    Weaver, AL
    Dale, JC
    Roberts, GD
    Henry, NK
    Ilstrup, DM
    Rosenblatt, JE
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1996, 34 (01) : 20 - 24
  • [3] COMPARISON OF LYSIS-CENTRIFUGATION WITH A BIPHASIC BLOOD CULTURE-MEDIUM FOR THE RECOVERY OF AEROBIC AND FACULTATIVELY ANAEROBIC-BACTERIA
    HENRY, NK
    GREWELL, CM
    VANGREVENHOF, PE
    ILSTRUP, DM
    WASHINGTON, JA
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1984, 20 (03) : 413 - 416
  • [4] CLINICAL COMPARISON OF ISOLATOR AND THIOL BROTH WITH ESP AEROBIC AND ANAEROBIC BOTTLES FOR RECOVERY OF PATHOGENS FROM BLOOD
    KELLOGG, JA
    BANKERT, DA
    MANZELLA, JP
    PARSEY, KS
    SCOTT, SL
    CAVANAUGH, SH
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1994, 32 (09) : 2050 - 2055
  • [5] CONTROLLED CLINICAL-EVALUATION OF ISOLATOR AND ESP AEROBIC BLOOD CULTURE SYSTEMS FOR DETECTION OF BLOOD-STREAM INFECTIONS
    KIRKLEY, BA
    EASLEY, KA
    WASHINGTON, JA
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1994, 32 (06) : 1547 - 1549
  • [6] MULTICENTER CLINICAL-EVALUATION OF A CONTINUOUS MONITORING BLOOD CULTURE SYSTEM USING FLUORESCENT-SENSOR TECHNOLOGY (BACTEC-9240)
    NOLTE, FS
    WILLIAMS, JM
    JERRIS, RC
    MORELLO, JA
    LEITCH, CD
    MATUSHEK, S
    SCHWABE, LD
    DORIGAN, F
    KOCKA, FE
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1993, 31 (03) : 552 - 557
  • [7] CONTROLLED CLINICAL COMPARISON OF ISOLATOR AND BACTEC-9240 AEROBIC-F RESIN BOTTLE FOR DETECTION OF BLOOD-STREAM INFECTIONS
    POHLMAN, JK
    KIRKLEY, BA
    EASLEY, KA
    WASHINGTON, JA
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1995, 33 (10) : 2525 - 2529