Rupture of the left main-stem bronchus by the tracheal portion of a double-lumen endobronchial tube

被引:31
作者
Sakuragi, T
Kumano, K
Yasumoto, M
Dan, K
机构
[1] Department of Anesthesiology, School of Medicine, Fukuoka University, Fukuoka
[2] Department of Anesthesiology, School of Medicine, Fukuoka University, Jonan-ku Fukuoka 814-01
关键词
complication; bronchial rupture; equipment; tubes; bronchial;
D O I
10.1111/j.1399-6576.1997.tb04869.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We report a rupture of the left main-stem bronchus following the insertion of a left-sided double-lumen endobronchial tube in a 76-yr-old woman with a short trachea. A fiberoptic bronchoscope was not used during the initial insertion of the tube and the depth of insertion resulted in approximately 5 cm in excess of the optimal level for this patient. The rupture had been caused by the tracheal portion of the double-lumen tube. This damage may have been avoided if a fiberoptic bronchoscope was used routinely as an introducer and for positioning of the endobronchial tube under direct vision.
引用
收藏
页码:1218 / 1220
页数:3
相关论文
共 10 条
[1]  
BRODSKY JB, 1991, ANESTH ANALG, V73, P570
[2]   RUPTURE OF THE LEFT MAIN-STEM BRONCHUS WITH A POLYVINYLCHLORIDE DOUBLE-LUMEN TUBE [J].
BURTON, NA ;
FALL, SM ;
LYONS, T ;
GRAEBER, GM .
CHEST, 1983, 83 (06) :928-929
[3]   RUPTURED BRONCHUS FOLLOWING ENDOBRONCHIAL INTUBATION - A CASE-REPORT [J].
FOSTER, JMG ;
LAU, OJ ;
ALIMO, EB .
BRITISH JOURNAL OF ANAESTHESIA, 1983, 55 (07) :687-688
[4]   TRACHEOBRONCHIAL RUPTURES DUE TO CUFFED CARLENS TUBES [J].
GUERNELLI, N ;
BRAGAGLIA, RB ;
BRICCOLI, A ;
MASTRORILLI, M ;
VECCHI, R .
ANNALS OF THORACIC SURGERY, 1979, 28 (01) :66-68
[5]   BRONCHIAL RUPTURE ASSOCIATED WITH THE USE OF A DOUBLE-LUMEN TUBE IN A SMALL ADULT [J].
HANNALLAH, M ;
GOMES, M .
ANESTHESIOLOGY, 1989, 71 (03) :457-459
[6]  
HANNALLAH MS, 1993, ANESTH ANALG, V77, P990
[7]   BRONCHIAL RUPTURE, A COMPLICATION OF USE OF THE ROBERTSHAW DOUBLE-LUMEN TUBE [J].
HEISER, M ;
STEINBERG, JJ ;
MACVAUGH, H ;
KLINEBERG, PL .
ANESTHESIOLOGY, 1979, 51 (01) :88-88
[8]  
OVASSAPIAN A, 1983, ANESTHESIOLOGY, V59, pA501
[9]   BRONCHOFIBERSCOPIC PLACEMENT OF A DOUBLE-LUMEN ENDOTRACHEAL-TUBE [J].
SHINNICK, JP ;
FREEDMAN, AP .
CRITICAL CARE MEDICINE, 1982, 10 (08) :544-545
[10]   TRACHEAL RUPTURE FOLLOWING THE INSERTION OF A DISPOSABLE DOUBLE-LUMEN ENDOTRACHEAL-TUBE [J].
WAGNER, DL ;
GAMMAGE, GW ;
WONG, ML .
ANESTHESIOLOGY, 1985, 63 (06) :698-700