A high-resolution susceptibility thermometer for use near 2 K

被引:6
作者
Baddar, H [1 ]
Fu, HY [1 ]
Larson, M [1 ]
Mulders, N [1 ]
Ahlers, G [1 ]
机构
[1] UNIV CALIF SANTA BARBARA, CTR NONLINEAR SCI, SANTA BARBARA, CA 93106 USA
关键词
D O I
10.1007/BF02570416
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
A relatively small and easily constructed high-resolution thermometer is described. It uses the magnetic salt copper ammonium bromide dihydrate (CAB). The temperature is read out using a de-SQUID. We determined the sensitivity of the thermometer as a function of temperature from 1.5 K to 4.3 K and for magnetic fields from 40 to 800 Gauss. Subnano-Kelvin resolution is achievable near 2 K. A mathematic model is used to understand the measured sensitivity and to improve the design.
引用
收藏
页码:2859 / 2860
页数:2