Prospective randomized comparison of irrigated-tip versus conventional-tip catheters for ablation of common flutter

被引:141
作者
Jaïs, P [1 ]
Shah, DC [1 ]
Haissaguerre, M [1 ]
Hocini, M [1 ]
Garrigue, S [1 ]
Le Metayer, P [1 ]
Clémenty, J [1 ]
机构
[1] Hop Cardiol Haut Leveque, F-33604 Bordeaux, France
关键词
catheter ablation; atrial flutter; catheters;
D O I
10.1161/01.CIR.101.7.772
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-Radiofrequency (RF) ablation of common flutter requires the creation of a complete ablation line to produce bidirectional conduction block in the cavotricuspid isthmus. An irrigated-tip ablation catheter has been shown to be effective in patients in whom conventional ablation has failed. This randomized study compares the efficacy and safety of this catheter with those of a conventional catheter for de novo flutter ablation. Methods and Results-Cavotricuspid ablation was performed with a conventional (n=26) or an irrigated-tip catheter (n=24). RF was applied for 60 minutes with a temperature-controlled mode: 65 degrees C to 70 degrees C up to 70 W with a conventional catheter or 50 degrees C up to 50 W (with a 17-mL/min saline flow rate) with the irrigated-tip catheter. The end point was the achievement of bidirectional isthmus block, and a crossover was performed after 21 unsuccessful applications. Procedural ablation parameters as well as number of applications, x-ray exposure, procedure duration, impedance rise, and clot formation were compared for each group. A coronary angiogram was performed before and after each ablation for the first 30 patients. Complete bidirectional isthmus block was achieved for all patients. Four patients crossed over from conventional to irrigated-tip catheters. The number of applications, procedure duration, and x-ray exposure were significantly higher with the conventional than with the irrigated-tip catheter: 13+/-10 versus 5+/-3 pulses, 53+/-41 versus 27+/-16 minutes, and 18+/-14 versus 9+/-6 minutes, respectively. No significant side effects occurred, and the coronary angiograms of the first 30 patients after ablation were unchanged. Conclusions-Irrigated-tip catheters were found to be more effective than and as safe as conventional catheters for flutter ablation, facilitating the rapid achievement of bidirectional isthmus block.
引用
收藏
页码:772 / 776
页数:5
相关论文
共 27 条
  • [1] Radiofrequency catheter ablation of common atrial flutter - Significance of palpitations and quality-of-life evaluation in patients with proven isthmus block
    Anselme, F
    Saoudi, N
    Poty, H
    Douillet, R
    Cribier, A
    [J]. CIRCULATION, 1999, 99 (04) : 534 - 540
  • [2] BELLET S, 1963, CLIN DISORDERS HEART, P144
  • [3] RADIOFREQUENCY CATHETER ABLATION OF ACCESSORY ATRIOVENTRICULAR CONNECTIONS IN 250 PATIENTS - ABBREVIATED THERAPEUTIC APPROACH TO WOLFF-PARKINSON-WHITE SYNDROME
    CALKINS, H
    LANGBERG, J
    SOUSA, J
    ELATASSI, R
    LEON, A
    KOU, W
    KALBFLEISCH, S
    MORADY, F
    [J]. CIRCULATION, 1992, 85 (04) : 1337 - 1346
  • [4] Electrophysiological effects of catheter ablation of inferior vena cava-tricuspid annulus isthmus in common atrial flutter
    Cauchemez, B
    Haissaguerre, M
    Fischer, B
    Thomas, O
    Clementy, J
    Coumel, P
    [J]. CIRCULATION, 1996, 93 (02) : 284 - 294
  • [5] RADIOFREQUENCY ABLATION OF THE INFERIOR VENA-CAVA - TRICUSPID-VALVE ISTHMUS IN COMMON ATRIAL-FLUTTER
    COSIO, FG
    LOPEZGIL, M
    GOICOLEA, A
    ARRIBAS, F
    BARROSO, JL
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1993, 71 (08) : 705 - 709
  • [6] RADIOFREQUENCY CATHETER ABLATION FOR THE TREATMENT OF HUMAN TYPE-1 ATRIAL-FLUTTER - IDENTIFICATION OF A CRITICAL ZONE IN THE REENTRANT CIRCUIT BY ENDOCARDIAL MAPPING TECHNIQUES
    FELD, GK
    FLECK, RP
    CHEN, PS
    BOYCE, K
    BAHNSON, TD
    STEIN, JB
    CALISI, CM
    IBARRA, M
    [J]. CIRCULATION, 1992, 86 (04) : 1233 - 1240
  • [7] Radiofrequency catheter ablation of common atrial flutter in 200 patients
    Fischer, B
    Jais, P
    Shah, DP
    Chouairi, S
    Haissaguerre, M
    Garrigues, S
    Poquet, F
    Gencel, L
    Clementy, J
    Marcus, FI
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1996, 7 (12) : 1225 - 1233
  • [8] High energy radiofrequency catheter ablation for common atrial flutter targeting the isthmus between the inferior vena cava and tricuspid valve annulus using a super long tip electrode
    Iesaka, Y
    Takahashi, A
    Goya, M
    Yamane, T
    Tokunaga, T
    Amemiya, H
    Fujiwara, H
    Nitta, J
    Nogami, A
    Aonuma, K
    Hiroe, M
    Marumo, F
    Hiraoka, M
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (02): : 401 - 409
  • [9] Irani WN, 1997, CIRCULATION, V95, P962
  • [10] CATHETER ABLATION OF ACCESSORY ATRIOVENTRICULAR PATHWAYS (WOLFF-PARKINSON-WHITE SYNDROME) BY RADIOFREQUENCY CURRENT
    JACKMAN, WM
    WANG, XZ
    FRIDAY, KJ
    ROMAN, CA
    MOULTON, KP
    BECKMAN, KJ
    MCCLELLAND, JH
    TWIDALE, N
    HAZLITT, HA
    PRIOR, MI
    MARGOLIS, PD
    CALAME, JD
    OVERHOLT, ED
    LAZZARA, R
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1991, 324 (23) : 1605 - 1611