Multiple hit readout of a microchannel plate detector with a three-layer delay-line anode

被引:304
作者
Jagutzki, O [1 ]
Cerezo, A
Czasch, A
Dörner, R
Hattass, M
Huang, M
Mergel, V
Spillmann, U
Ullmann-Pfleger, K
Weber, T
Schmidt-Böcking, H
Smith, GDW
机构
[1] Univ Frankfurt, Inst Kernphys, D-6000 Frankfurt, Germany
[2] RoentDek GmbH, Kelkheim, Germany
[3] Univ Oxford, Dept Mat, Oxford OX1 3PH, England
[4] Oxford Nanosci Ltd, Oxford, England
关键词
delay line; fragment imaging; imaging; microchannel plate; multihit; three-dimensional (3-D) detection;
D O I
10.1109/TNS.2002.803889
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We have developed a delay-line readout technique for microchannel plate detectors with an increased acceptance for multiple hit events compared to standard two-layer delay-line anodes. This technique allows unambiguous determination of arrival time and position of at least two simultaneously detected particles, and/or to detect an even larger number of particles in a shower, as long as any two particles do not arrive both at the same time and at the same position within certain limits. We demonstrate and discuss the abilities and limitations of this technique and the relevance for certain experimental tasks.
引用
收藏
页码:2477 / 2483
页数:7
相关论文
共 29 条
[1]   A SYSTEM FOR CORRELATED FRAGMENT DETECTION IN DISSOCIATION EXPERIMENTS [J].
BECKER, J ;
BECKORD, K ;
WERNER, U ;
LUTZ, HO .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 337 (2-3) :409-415
[2]   A NEW TYPE OF MULTIPARTICLE IMAGING DETECTOR [J].
BELKACEM, A ;
FAIBIS, A ;
KANTER, EP ;
KOENIG, W ;
MITCHELL, RE ;
VAGER, Z ;
ZABRANSKY, BJ .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (03) :945-952
[3]   Performance of an energy-compensated three-dimensional atom probe [J].
Cerezo, A ;
Godfrey, TJ ;
Sijbrandij, SJ ;
Smith, GDW ;
Warren, PJ .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1998, 69 (01) :49-58
[4]   Cold Target Recoil Ion Momentum Spectroscopy:: a 'momentum microscope' to view atomic collision dynamics [J].
Dörner, R ;
Mergel, V ;
Jagutzki, O ;
Spielberger, L ;
Ullrich, J ;
Moshammer, R ;
Schmidt-Böcking, H .
PHYSICS REPORTS-REVIEW SECTION OF PHYSICS LETTERS, 2000, 330 (2-3) :95-192
[5]   SIMPLE 2-DIMENSIONAL POSITION-SENSITIVE DETECTOR WITH SHORT DEAD-TIME FOR COINCIDENCE EXPERIMENTS [J].
ELAND, JHD .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1994, 5 (12) :1501-1504
[6]   MUPPATS - A MULTIPARTICLE 3D IMAGING DETECTOR SYSTEM [J].
FAIBIS, A ;
KOENIG, W ;
KANTER, EP ;
VAGER, Z .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1986, 13 (1-3) :673-677
[8]   Multilayer anode with crossed serpentine delay lines for high spatial resolution readout of microchannel plate detectors [J].
Friedman, PG ;
Cuza, RA ;
Fleischman, JR ;
Martin, C ;
Schiminovich, D ;
Doyle, DJ .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1996, 67 (02) :596-608
[9]   Fast position and time resolved read-out of micro-channelplates with the delay-line technique for single particle and photon detection [J].
Jagutzki, O ;
Mergel, V ;
Ullmann-Pfleger, K ;
Spielberger, L ;
Meyer, U ;
Dorner, R ;
Schmidt-Bocking, H .
IMAGING SPECTROMETRY IV, 1998, 3438 :322-333
[10]   A broad-application microchannel-plate detector system for advanced particle or photon detection tasks:: large area imaging, precise multi-hit timing information and high detection rate [J].
Jagutzki, O ;
Mergel, V ;
Ullmann-Pfleger, K ;
Spielberger, L ;
Spillmann, U ;
Dörner, R ;
Schmidt-Böcking, H .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 477 (1-3) :244-249