Quality of mechanical, manual standard and active compression-decompression CPR on the arrest site and during transport in a manikin model

被引:95
作者
Sunde, K
Wik, L
Steen, PA
机构
[1] Ullevål University Hospital, Inst. for Exp. Medical Research
[2] Norwegian Air Ambulance, Dept. Res. and Educ. in Acute Med.
[3] Ullevål University Hospital, Department of Anesthesiology
关键词
cardiopulmonary resuscitation; transport; external chest compressions; active-compression-decompressions; mechanical; manikin;
D O I
10.1016/S0300-9572(96)01087-8
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
The quality of mechanical CPR (M-CPR) was compared with manual standard CPR (S-CPR) and active compression-decompression CPR (ACD-CPR) performed by paramedics on the site of a cardiac arrest and during manual and ambulance transport. Each technique was performed 12 times on manikins using teams from a group of 12 paramedic students with good clinical CPR experience using a random cross-over design. Except for some lost ventilations the CPR effort using the mechanical device adhered to the European Resuscitation Council guidelines, with an added time requirement of median 40 s for attaching the device compared with manual standard CPR. Throughout the study, in comparison with mechanical CPR the quality of CPR with either manual method was significantly worse. In particular, there were considerable individual variations during stretcher transport. With S-CPR and ACD-CPR the median compression times were 38 and 31%, significantly lower than the recommended 50%, and 46-98% of the decompression efforts with ACD-CPR were too weak, particularly during transport on the stairs. With both manual methods, there were no significant differences in the CPR effort between the site of the arrest and the ambulance transport. However, compression rates were reduced and became more erratic during stretcher transport to the ambulance. When walking horizontally, a median of 19% of S-CPR compressions and 84% of ACD-CPR compressions were to weak. On the stairs, 68% of S-CPR compressions and 100% of ACD-CPR compressions were too weak. In conclusion, when evaluated on a manikin, in comparison with manual standard and ACD-CPR, mechanical CPR adhered more closely to ERC guidelines. This was particularly true when performing CPR during transport on a stretcher. (C) 1997 Elsevier Science Ireland Ltd.
引用
收藏
页码:235 / 242
页数:8
相关论文
共 28 条
[1]   Bystander-initiated cardiopulmonary resuscitation out-of-hospital. A first description of the bystanders and their experiences [J].
Axelsson, A ;
Herlitz, J ;
Ekstrom, L ;
Holmberg, S .
RESUSCITATION, 1996, 33 (01) :3-11
[2]   RELATIONSHIP OF BLOOD-PRESSURE AND FLOW DURING CPR TO CHEST COMPRESSION AMPLITUDE - EVIDENCE FOR AN EFFECTIVE COMPRESSION THRESHOLD [J].
BABBS, CF ;
VOORHEES, WD ;
FITZGERALD, KR ;
HOLMES, HR ;
GEDDES, LA .
ANNALS OF EMERGENCY MEDICINE, 1983, 12 (09) :527-532
[3]   Guidelines for the basic management of the airway and ventilation during resuscitation - A statement by the Airway and Ventilation Management Working Group of the European Resuscitation Council [J].
Baskett, PJF ;
Bossaert, L ;
Carli, P ;
Chamberlain, D ;
Dick, W ;
Nolan, JP ;
Parr, MJA ;
Scheidegger, D ;
Zideman, D ;
Blancke, W ;
Delooz, H ;
Handley, A ;
Kettler, D ;
Kloeck, W ;
Kramer, E ;
Quan, L ;
Studer, W ;
VanDrenth, A .
RESUSCITATION, 1996, 31 (03) :187-200
[4]   DISTINCT CRITERIA FOR TERMINATION OF RESUSCITATION IN THE OUT-OF-HOSPITAL SETTING [J].
BONNIN, MJ ;
PEPE, PE ;
KIMBALL, KT ;
CLARK, PS .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1993, 270 (12) :1457-1462
[5]   ACTIVE COMPRESSION-DECOMPRESSION CPR IMPROVES VITAL ORGAN PERFUSION IN A DOG-MODEL OF VENTRICULAR-FIBRILLATION [J].
CHANG, MW ;
COFFEEN, P ;
LURIE, KG ;
SHULTZ, J ;
BACHE, RJ ;
WHITE, CW .
CHEST, 1994, 106 (04) :1250-1259
[6]   IMPROVED BLOOD-FLOW DURING PROLONGED CARDIOPULMONARY-RESUSCITATION WITH 30-PERCENT DUTY CYCLE IN INFANT PIGS [J].
DEAN, JM ;
KOEHLER, RC ;
SCHLEIEN, CL ;
ATCHISON, D ;
GERVAIS, H ;
BERKOWITZ, I ;
TRAYSTMAN, RJ .
CIRCULATION, 1991, 84 (02) :896-904
[7]   CARDIAC-OUTPUT DURING CARDIOPULMONARY RESUSCITATION AT VARIOUS COMPRESSION RATES AND DURATIONS [J].
FITZGERALD, KR ;
BABBS, CF ;
FRISSORA, HA ;
DAVIS, RW ;
SILVER, DI .
AMERICAN JOURNAL OF PHYSIOLOGY, 1981, 241 (03) :H442-H448
[8]   INTRATHORACIC PRESSURE-FLUCTUATIONS MOVE BLOOD DURING CPR - COMPARISON OF HEMODYNAMIC DATA WITH PREDICTIONS FROM A MATHEMATICAL-MODEL [J].
HALPERIN, HR ;
TSITLIK, JE ;
BEYAR, R ;
CHANDRA, N ;
GUERCI, AD .
ANNALS OF BIOMEDICAL ENGINEERING, 1987, 15 (3-4) :385-403
[9]   The relationship between rate of chest compression and compression:relaxation ratio [J].
Handley, AJ ;
Handley, JA .
RESUSCITATION, 1995, 30 (03) :237-241
[10]   GUIDELINES FOR BASIC LIFE-SUPPORT [J].
HOLMBERG, S ;
HANDLEY, A ;
BAHR, J ;
BASKETT, P ;
BOSSAERT, L ;
CHAMBERLAIN, D ;
DICK, W ;
VANDRENTH, A ;
EKSTROM, L ;
JUCHEMS, R ;
KETTLER, D ;
MARSDEN, A ;
MOESCHLER, O ;
MONSIEURS, K ;
PETIT, P ;
PRIMOZ, H ;
SCHUTTLER, J .
RESUSCITATION, 1992, 24 (02) :103-110