Photometric measurement of trace As(III) and As(V) in drinking water

被引:52
作者
Dasgupta, PK [1 ]
Huang, HL
Zhang, GF
Cobb, GP
机构
[1] Texas Tech Univ, Dept Chem & Biochem, Lubbock, TX 79409 USA
[2] Texas Tech Univ, Inst Environm Hlth, Lubbock, TX 79409 USA
关键词
light-emitting diode; photometric method; phosphate; flow injection analysis;
D O I
10.1016/S0039-9140(02)00264-3
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple, fast and sensitive light-emitting diode (LED)-based photometric method for the differential determination of ppb-ppm levels of As(III) and As(V) in potable water in the presence of ppm levels of phosphate was developed. The detection chemistry is based on the well-known formation of arsenomolybdate, followed by reduction to heteropoly blue. The front-end of the measurement system is configured to selectively retain P(V) and As(V), based on the considerable difference of the pK(a) of the corresponding acids relative to As(III). Thus, it is As(III) that is injected into the medium, oxidized in-line with KBrO3 to As(V) and forms Mo-blue that is detected by an LED-based detector. Only As(III) is measured if the sample is injected as such; if all As in the sample is prereduced to As(III) (by the addition of cysteine, in a provided in-line arrangement), the system measures As(V) + As(III). In the present form, limit of detection (LOD) (S/N= 3) is less than 8 mug l(-1) As, and the linear range extends to 2.4 mg l(-1). Potential interference from dissolved silica and Fe(III) is eliminated by the addition of NaF to the sample. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:153 / 164
页数:12
相关论文
共 46 条