Deficiencies of active electronic radiation protection dosemeters in pulsed fields

被引:56
作者
Ankerhold, U. [1 ]
Hupe, O. [1 ]
Ambrosi, P. [1 ]
机构
[1] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
关键词
D O I
10.1093/rpd/ncp099
中图分类号
X [环境科学、安全科学];
学科分类号
083001 [环境科学];
摘要
Nowadays nearly all radiation fields used for X-ray diagnostics are pulsed. These fields are characterised by a high dose rate during the pulse and a short pulse duration in the range of a few milliseconds. The use of active electronic dosemeters has increased in the past few years, but these types of dosemeters might possibly not measure reliably in pulsed radiation fields. Not only personal dosemeters but also area dosemeters that are used mainly for dose rate measurements are concerned. These cannot be substituted by using passive dosemeter types. The characteristics of active electronic dosemeters determined in a continuous radiation field cannot be transferred to those in pulsed fields. Some provisional measurements with typical electronic dosemeters in pulsed radiation fields are presented to reveal this basic problem.
引用
收藏
页码:149 / 153
页数:5
相关论文
共 11 条
[1]
A prototype ionisation chamber as a secondary standard for the measurement of personal dose equivalent, Hp(10), on a slab phantom [J].
Ankerhold, U ;
Behrens, R ;
Ambrosi, P .
RADIATION PROTECTION DOSIMETRY, 1999, 86 (03) :167-173
[2]
ANKERHOLD U, 2000, PTBDOS34
[3]
[Anonymous], 61267 IEC
[4]
[Anonymous], 1982, 34 ICRU
[5]
*BUND UMW, 2007, NAT REAKT UMW STRAHL
[6]
Intercomparison of active personal dosemeters in interventional radiology [J].
Clairand, Isabelle ;
Struelens, Lara ;
Bordy, Jean-Marc ;
Daures, Josiane ;
Debroas, Jacques ;
Denozieres, Marc ;
Donadille, Laurent ;
Gouriou, Jean ;
Itie, Christian ;
Vaz, Pedro ;
d'Errico, Francesco .
RADIATION PROTECTION DOSIMETRY, 2008, 129 (1-3) :340-345
[7]
*INT AT EN AG, 2007, 1564 IAEATECDOC
[8]
*INT EL COMM, 2008, 61526 IECCDV
[9]
*INT EL COMM, 2005, 61005 IEC
[10]
*INT EL COMM, 2009, 608461 IECFDIS