Estimation of the Non-Measurable State Variables of a Transcutaneous Energy Transmission System for Artificial Human Implants Using Extended Kalman Filters

被引:5
作者
Liu, Wei [1 ]
Tang, Houjun [1 ]
Fang, Wan [1 ]
Ye, Pengsheng [1 ]
机构
[1] Shanghai Jiao Tong Univ, Dept Elect Engn, Shanghai, Peoples R China
关键词
Non-measurable state variables; Transcutaneous energy transmission system; Artificial human implants; Extended Kalman filter; Estimator;
D O I
10.1007/s00034-009-9099-2
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
080906 [电磁信息功能材料与结构]; 082806 [农业信息与电气工程];
摘要
Due to the separation of the two sides of the coupling network, the acquisition of data on the operating state variables of a transcutaneous energy transmission system (TETS) inside the human body is difficult. A non-measurable state estimation approach is used in this work to facilitate the estimation of non-measurable variables on the secondary side of the TETS, including the current of the secondary coil and the output voltage of the secondary side. The estimation algorithm is based on a discrete dynamic mathematical model of the TETS. Following this model, using the extended Kalman filter (EKF) algorithm, the complexity of the TETS for artificial human implants can be reduced, while the reliability is simultaneously enhanced. Additionally, as an adaptive filter, the EKF can also successfully filter out processing noise during energy transmission. All of the results are verified by simulation using MATLAB.
引用
收藏
页码:581 / 593
页数:13
相关论文
共 12 条
[1]
Anderson B.D., 2012, Optimal Filtering
[2]
Examination of circuit parameters for stable high efficiency TETS for artificial hearts [J].
Arai, S ;
Miura, H ;
Sato, F ;
Matsuki, H ;
Sato, T .
IEEE TRANSACTIONS ON MAGNETICS, 2005, 41 (10) :4170-4172
[3]
Modeling and analyzing an inductive contactless power transfer system for artificial hearts using the generalized state space averaging method [J].
Fang, Wan ;
Tang, Houjun ;
Liu, Wei .
JOURNAL OF COMPUTATIONAL AND THEORETICAL NANOSCIENCE, 2007, 4 (7-8) :1412-1416
[4]
FRANKLIN GF, 1998, DIGITAL CONTROL DYNA, P388
[5]
A wide-band power-efficient inductive wireless link for implantable microelectronic devices using multiple-carriers [J].
Ghovanloo, Maysam ;
Atluri, Suresh .
IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS I-REGULAR PAPERS, 2007, 54 (10) :2211-2221
[6]
Liu W, 2007, C IND ELECT APPL, P1873
[7]
Improvement of the transcutaneous energy transmission system utilizing ferrite cored coils for artificial hearts [J].
Miura, Hidekazu ;
Arai, Shinsuke ;
Kakubari, Yasuyuki ;
Sato, Fumihiro ;
Matsuki, Hidetoshi ;
Sato, Tadakuni .
IEEE TRANSACTIONS ON MAGNETICS, 2006, 42 (10) :3578-3580
[8]
Generalized Averaging Method for Power Conversion Circuits [J].
Sanders, Seth R. ;
Noworolski, J. Mark ;
Liu, Xiaojun Z. ;
Verghese, George C. .
IEEE TRANSACTIONS ON POWER ELECTRONICS, 1991, 6 (02) :251-259
[9]
A new contactless power-signal transmission device for implanted functional electrical stimulation [J].
Sato, F ;
Nomoto, T ;
Kano, G ;
Matsuki, H ;
Sato, T .
IEEE TRANSACTIONS ON MAGNETICS, 2004, 40 (04) :2964-2966
[10]
Powering an artificial heart: Birth of the inductively coupled-radio frequency system in 1960 [J].
Schuder, JC .
ARTIFICIAL ORGANS, 2002, 26 (11) :909-915