Mode switching anomalies: A patient who remained symptomatic during paroxysmal atrial tachyarrhythmias despite a mode switching pacemaker

被引:7
作者
Giudici, MC
Orias, DW
机构
[1] Division of Cardiology, Genesis Medical Center, Davenport, IA
[2] Genesis Medical Center, Davenport, IA 52803
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1997年 / 20卷 / 07期
关键词
mode switching; rate smoothing; atrial tachyarrhythmias;
D O I
10.1111/j.1540-8159.1997.tb03583.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A patient with an automatic mode switching pacemaker continued to experience discomfort during the onset of poroxysmal supraventricular tachyarrhythmias. investigation revealed that the patient was sensing abrupt rate changes as the ventricular paced rate tracked the tachycardia during onset and detection phases. Her pacemaker was replaced with a new device with both mode switching and rate smoothing capabilities with resultant elimination of symptoms.
引用
收藏
页码:1883 / 1884
页数:2
相关论文
共 3 条
[1]   ANTIARRHYTHMIC DRUG-THERAPY AND CARDIAC MORTALITY IN ATRIAL-FIBRILLATION [J].
FLAKER, GC ;
BLACKSHEAR, JL ;
MCBRIDE, R ;
KRONMAL, RA ;
HALPERIN, JL ;
HART, RG .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 20 (03) :527-532
[2]   CLINICAL INTEREST OF A SENSOR DRIVEN ALGORITHM LIMITING VENTRICULAR PACING RATE DURING SUPRAVENTRICULAR TACHYCARDIA IN DUAL CHAMBER PACING [J].
MAHAUX, V ;
VERBOVEN, Y ;
WALEFFE, A ;
KULBERTUS, H .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (11) :1862-1866
[3]   THE AUTOMATIC-MODE SWITCH FUNCTION IN SUCCESSIVE GENERATIONS OF MINUTE VENTILATION SENSING DUAL-CHAMBER RATE-RESPONSIVE PACEMAKERS [J].
PROVENIER, F ;
JORDAENS, L ;
VERSTRAETEN, T ;
CLEMENT, DL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (11) :1913-1919