Average and peak contact stress distribution evaluation of total knee arthroplasties

被引:93
作者
Szivek, JA
Anderson, PL
Benjamin, JB
机构
[1] Orthopedic Research Laboratory, Department of Surgery, University of Arizona, Tucson, AZ
[2] Orthopedic Research Laboratory, Arizona Health Sciences Center, University of Arizona, Tucson
关键词
artificial knee; contact geometry; tibiofemoral contact stress; patellofemoral contact stress; Fuji film;
D O I
10.1016/S0883-5403(96)80137-9
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Seven total knee arthroplasty systems were tested to determine contact stress patterns and contact areas using a calibrated Fuji film stress analysis technique. Knees were loaded to 2,000 N (204 kg) at 15 degrees, 60 degrees, 90 degrees, and 135 degrees flexion at 24 and 37 degrees C. Evaluation of stresses al 37 degrees C al 15 degrees and 60 degrees using an average contact stress assessment technique indicated that the LCS meniscal bearing knee system, (DePuy, Warsaw, IN), the AMK knee with a constrained insert (DePuy), and the PFC knee with a posterior-lipped insert (Johnson and Johnson, Raynham, MA) had the lowest average contact stresses (near or below 10 MPa). The PFC with a regular insert (Johnson and Johnson), the Ortholoc II (Dow Corning Wright, Arlington, TN), and the AMK with a regular insert (DePuy) had intermediate contact stresses. The AMK with a Hylamer-M insert (DePuy) and the MG II (Zimmer, Warsaw, IN) had the highest average contact stresses (near or above 20 MPa). A stress-calibrated Fuji film measurement technique has shown that an assessment of ranges of contact stress provides much more information about regions of expected wear than an assessment of average contact stresses. Testing of the tibiofemoral articulation of artificial knees revealed that all knees had some areas of contact with maximum stresses in excess of 15 MPa. As the yield strength of ultrahigh-molecular-weight polyethylene is approximately 15 MPa, all tibial inserts could wear to some extent. Peak contact stresses at four test angles of the AMK, Series 7000 (Osteonics, Allendale, NJ:) Genesis (Smith & Nephew Orthopaedics, Memphis, TN), and MG II patellofemoral articulations were high (above 30 MPa). Contact areas varied from line-shaped to bilateral circular or elliptical shapes. The LCS knee system experienced substantially lower patellofemoral contact stresses and larger contact areas. Changes in conformity of knee designs are warranted to overcome wear problems. Peak contact stresses measured from the LCS meniscal bearing tibiofemoral and patellofemoral joint were in excess of 30 MPa in some areas at low flexion angles. This design does create large areas of contact at very low contact pressures, however, and for this reason is expected to wear less than other designs.
引用
收藏
页码:952 / 963
页数:12
相关论文
共 24 条
[1]
THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[2]
THE EFFECT OF CONFORMITY AND PLASTIC THICKNESS ON CONTACT STRESSES IN METAL-BACKED PLASTIC IMPLANTS [J].
BARTEL, DL ;
BURSTEIN, AH ;
TODA, MD ;
EDWARDS, DL .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1985, 107 (03) :193-199
[3]
BLUNN GW, 1991, CLIN ORTHOPAEDICS, V273, P253
[4]
Buechel F F, 1991, Orthop Rev, V20, P50
[5]
CAMERON HU, 1993, CAN J SURG, V36, P560
[6]
COLLIER JP, 1991, CLIN ORTHOP RELAT R, V273, P198
[7]
COLLIER JP, 1991, CLIN ORTHOP RELAT R, V273, P232
[8]
A QUANTITATIVE TECHNIQUE FOR REPORTING SURFACE DEGRADATION PATTERNS OF UHMWPE COMPONENTS OF RETRIEVED TOTAL KNEE REPLACEMENTS [J].
CORNWALL, GB ;
BRYANT, JT ;
HANSSON, CM ;
RUDAN, J ;
KENNEDY, LA ;
COOKE, TDV .
JOURNAL OF APPLIED BIOMATERIALS, 1995, 6 (01) :9-18
[9]
CROWNINSHIELD RD, 1982, T ORTHOP RES SOC, V7, P127
[10]
DAVIDSON JA, 1987, J BIOMED MATER RES-A, V21, P261