Validation of the Microlife BPA100 Plus device for self-home blood pressure measurement according to the International Protocol

被引:27
作者
Stergiou, George S. [1 ]
Giovas, Periklis R. [1 ]
Neofytou, Michalis S. [1 ]
Adamopoulos, Dionysios N. [1 ]
机构
[1] Univ Athens, Dept Med 3, Sotiria Hosp, Hypertens Ctr, Athens 11527, Greece
关键词
accuracy; European Society of Hypertension; home blood pressure; International Protocol; Microlife; self-measurement; validation;
D O I
10.1097/01.mbp.0000209071.84965.bf
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective: Current guidelines recommend that only validated devices for blood pressure measurement should be used. This study presents the validation results of the Microlife BPA100 Plus monitor for self-home blood pressure measurement, according to the European Society of Hypertension International Protocol. Methods: Fifteen study participants were included in phase 1 and an additional 18 in phase 2 (total 33). Simultaneous blood pressure measurements were taken by two trained observers (Y tube connected mercury sphygmomanometers) four times, sequentially with three measurements taken using the tested device. Absolute differences between observer and device blood pressure were classified into three zones (within 5, 10 and 15 mmHg). The number of readings with a difference within 5 mmHg was calculated for each individual. Results: In phase 1, the device produced 32, 42 and 43 measurements within 5, 10 and 15 mmHg, respectively, for systolic blood pressure and 31, 45 and 45 for diastolic blood pressure. In phase 2.1, the device produced 71, 87 and 96 measurements within 5, 10 and 15 mmHg, respectively, for systolic blood pressure, and 71, 98 and 99 measurements for diastolic blood pressure. In phase 2.2, 26 participants had at least two of their differences within 5 mmHg and three participants had no differences within 5 mmHg for systolic blood pressure, whereas for diastolic blood pressure 24 and three participants, respectively. Mean systolic blood pressure differences were -2.0 +/- 6.0 mmHg and diastolic blood pressure differences were 3.1 +/- 4.1 mmHg. Conclusions: The Microlife BPA100 Plus device for self-home blood pressure measurement comfortably passes the validation requirements of the International Protocol and therefore can be recommended for clinical use in the adult population.
引用
收藏
页码:157 / 160
页数:4
相关论文
共 9 条
[1]  
Association for the Advancement of Medical Instrumentation, 1993, AM NAT STAND EL AUT
[2]   Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure [J].
Chobanian, AV ;
Bakris, GL ;
Black, HR ;
Cushman, WC ;
Green, LA ;
Izzo, JL ;
Jones, DW ;
Materson, BJ ;
Oparil, S ;
Wright, JT ;
Roccella, EJ .
HYPERTENSION, 2003, 42 (06) :1206-1252
[3]  
Hemmelgarn Brenda R, 2005, Can J Cardiol, V21, P645
[4]   2003 European society of hypertension -: European Society of Cardiology guidelines for the management of arterial hypertension [J].
Mancia, G ;
Rosei, EA ;
Cifkova, R ;
DeBacker, G ;
Erdine, S ;
Fagard, R ;
Farsang, C ;
Heagerty, AM ;
Kawecka-Jaszcs, K ;
Kiowski, W ;
Kjeldsen, S ;
Lüscher, T ;
McInnes, G ;
Mallion, JM ;
Brien, EO ;
Poulter, NR ;
Priori, SG ;
Rahn, KH ;
Rodicio, JL ;
Ruilope, LM ;
Safar, M ;
Staessen, JA ;
van Zwieten, P ;
Waeber, B ;
Williams, B ;
Zanchetti, A ;
Zannad, F .
JOURNAL OF HYPERTENSION, 2003, 21 (06) :1011-1053
[5]   European Society of hypertension recommendations for conventional, ambulatory and home blood pressure measurement [J].
O'Brien, E ;
Asmar, R ;
Beilin, L ;
Imai, Y ;
Mallion, JM ;
Mancia, G ;
Mengden, T ;
Myers, M ;
Padfield, P ;
Palatini, P ;
Parati, G ;
Pickering, T ;
Redon, J ;
Staessen, J ;
Stergiou, G ;
Verdecchia, P .
JOURNAL OF HYPERTENSION, 2003, 21 (05) :821-848
[6]   Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults [J].
O'Brien, E ;
Pickering, T ;
Asmar, R ;
Myers, M ;
Parati, G ;
Staessen, J ;
Mengden, T ;
Imai, Y ;
Waeber, B ;
Palatini, P ;
Atkins, N ;
Gerin, W .
BLOOD PRESSURE MONITORING, 2002, 7 (01) :3-17
[7]  
O'Brien E., 1993, J HYPERTENSION S2, V11, P43, DOI DOI 10.1097/00004872-199306000-00013
[8]   Self monitoring of blood pressure at home - Is an important adjunct to clinic measurements [J].
Stergiou, G ;
Mengden, T ;
Padfield, PL ;
Parati, G ;
O'Brien, E .
BMJ-BRITISH MEDICAL JOURNAL, 2004, 329 (7471) :870-871
[9]  
2005, ED TRUST DEVICES BLO