Rapid multiexcitation fluorescence spectroscopy system for in vivo tissue diagnosis

被引:92
作者
Zangaro, RA
Silveira, L
Manoharan, R
Zonios, G
Itzkan, I
Dasari, RR
VanDam, J
Feld, MS
机构
[1] MIT,GEORGE R HARRISON SPECT LAB,CAMBRIDGE,MA 02139
[2] BRIGHAM & WOMENS HOSP,DIV GASTROENTEROL,BOSTON,MA 02115
来源
APPLIED OPTICS | 1996年 / 35卷 / 25期
关键词
instrumentation; fluorescence spectroscopy; rapid excitation-emission matrix; human tissue;
D O I
10.1364/AO.35.005211
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We have designed, fabricated, and tested a compact, transportable, excitation-emission spectrofluorimeter with optical-fiber light delivery and collection for use in rapid analysis of tissues in a clinical setting. This system provides up to eleven different excitation wavelengths, permitting collection of all the corresponding emission spectra in approximately 600 ms. It uses a N-2 laser that pumps a sequence of dyes placed in cuvettes on a rotating wheel. A white-light excitation source permits acquisition of the tissue's diffuse reflectance spectrum on each cycle. Return fluorescence and reflected light are dispersed by a small spectrograph and detected by a photodiode-array detector. The system can celled a single-shot spectrum from biological tissue with a signal-to-noise ratio in excess of 50:1. (C) 1996 Optical Society of America
引用
收藏
页码:5211 / 5219
页数:9
相关论文
共 27 条
[1]  
[Anonymous], PHOTOMULTIPLIER TUBE
[2]  
BIGIO IJ, 1994, P SOC PHOTO-OPT INS, V2324, P46
[3]   PORTABLE LASER SPECTROFLUOROMETER SYSTEM FOR IN-VIVO HUMAN TISSUE FLUORESCENCE STUDIES [J].
BRENNAN, JF ;
ZONIOS, GI ;
WANG, TD ;
RAVA, RP ;
HAYES, GB ;
DASARI, RR ;
FELD, MS .
APPLIED SPECTROSCOPY, 1993, 47 (12) :2081-2086
[4]  
Cothren R. M., 1986, Lasers in the Life Sciences, V1, P1
[5]   NEW INSTRUMENTATION FOR USE IN EXCITATION-EMISSION FLUORESCENCE-POLARIZATION MEASUREMENTS [J].
DESTRAMPE, KA ;
HIEFTJE, GM .
APPLIED SPECTROSCOPY, 1993, 47 (10) :1548-1554
[6]  
DUARTE FJ, 1990, DYE LASER PRINCIPLES, P17
[7]  
*EXC CHEM CO, 1995, EXC CHEM CO CAT
[8]  
FORRER M, 1995, P SOC PHOTO-OPT INS, V2324, P84
[9]   FLUORESCENCE SPECTROSCOPY OF TURBID MEDIA - AUTOFLUORESCENCE OF THE HUMAN AORTA [J].
KEIJZER, M ;
RICHARDSKORTUM, RR ;
JACQUES, SL ;
FELD, MS .
APPLIED OPTICS, 1989, 28 (20) :4286-4292
[10]   Spectroscopic diagnosis of bladder cancer with elastic light scattering [J].
Mourant, JR ;
Bigio, IJ ;
Boyer, J ;
Conn, RL ;
Johnson, T ;
Shimada, T .
LASERS IN SURGERY AND MEDICINE, 1995, 17 (04) :350-357