Damage to the laryngeal mask by residual fluid in the cuff

被引:8
作者
Asai, T [1 ]
Koga, K [1 ]
Morris, S [1 ]
机构
[1] LLANDOUGH HOSP, NHS TRUST, PENARTH CF64 2XX, S GLAM, WALES
关键词
equipment; laryngeal mask airway;
D O I
10.1111/j.1365-2044.1997.242-az0377.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
It has been suggested that, in some situations, the cuff of the laryngeal mask should be filled with fluid. We speculated that this practice might damage the device during sterilisation in an autoclave. We studied whether injection of a small volume of water into the cuff of the laryngeal mask and subsequent sterilisation affected the integrity of the cuff. First, a pressure-volume relationship for each of 20 new masks was obtained by inflating the cuff with increasing volumes of air (5-45 mi). The masks were then randomly allocated into four groups (W-0, W-0.25, W-0.5 or W-1.0), 0, 0.25, 0.5 or 1.0 mi of water was injected into the cuff and the mask was then sterilised in an autoclave. After sterilisation, the shape of the cuff was examined and pressure-volume relationships were obtained again. The baseline intracuff pressures were similar in the four groups. After sterilisation, the pressure was significantly lower in groups W-0.25, W-0.5 and W-1.0 than in group W-0 (p < 0.05). Two masks in group W-1.0 lost their integrity, resulting in herniation of and rupture of the cuff. We conclude that the cuff of the laryngeal mask should not be inflated with fluid unless the device is discarded afterwards.
引用
收藏
页码:977 / 981
页数:5
相关论文
共 15 条
[1]   MISUSE OF THE LARYNGEAL MASK AIRWAY [J].
ASAI, T ;
VAUGHAN, RS .
ANAESTHESIA, 1994, 49 (06) :467-469
[2]   INFLATION OF THE CUFF OF THE LARYNGEAL MASK [J].
ASAI, T ;
MORRIS, S .
ANAESTHESIA, 1994, 49 (12) :1098-1099
[3]   DAMAGE TO LARYNGEAL MASKS DURING STERILIZATION [J].
BIRO, P .
ANESTHESIA AND ANALGESIA, 1993, 77 (05) :1079-1079
[4]   Optimal shape of the laryngeal mask cuff - The influence of three deflation techniques [J].
Brimacombe, J ;
Brain, A ;
Branagan, H ;
Spry, M ;
Schofield, J .
ANAESTHESIA, 1996, 51 (07) :673-676
[5]   SPONTANEOUS REINFLATION CHARACTERISTICS OF THE LARYNGEAL MASK AIRWAY [J].
BRIMACOMBE, J .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1994, 41 (09) :873-873
[6]   LARYNGEAL MASK RESIDUAL VOLUME AND DAMAGE DURING STERILIZATION [J].
BRIMACOMBE, JR .
ANESTHESIA AND ANALGESIA, 1994, 79 (02) :391-391
[7]   ENDOBRONCHIAL CUFF PRESSURES [J].
COBLEY, M ;
KIDD, JF ;
WILLIS, BA ;
VAUGHAN, RS .
BRITISH JOURNAL OF ANAESTHESIA, 1993, 70 (05) :576-578
[8]  
CRAWFORD M, 1992, ANAESTHESIA, V47, P76, DOI 10.1111/j.1365-2044.1992.tb01975.x
[9]  
Handsworth JL, 1996, ANAESTH INTENS CARE, V24, P728
[10]  
*INTR RES LTD, LARYNG MASK AIRW INS