Tidal volume and respiratory timing derived from a portable ventilation monitor

被引:32
作者
McCool, FD
Wang, J
Ebi, KL
机构
[1] Mem Hosp Rhode Isl, Dept Med, Pawtucket, RI 02809 USA
[2] Brown Univ, Sch Med, Pawtucket, RI USA
关键词
pulmonary ventilation; thorax; tidal volume;
D O I
10.1378/chest.122.2.684
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Study objectives: To determine the accuracy, of a portable magnetometer designed to measure tidal volume (VT), inspiratory time (TI), and expiratory, time (TE). Participants: Fourteen healthy subjects. Design: Subjects breathed over a sixfold range of VTS while at rest (sitting and standing) and during treadmill exercise. We then compared VT, Ti, and TE, measured by magnetometry (VTmag, TImag, and TEmag) with VT, Ti, and TE measured by spirometry (VTspiro, TIspiro, and TEspiro), respectively). Setting: Pulmonary function and exercise physiology, laboratories. Measurements and results: The sternal-umbilical distance and the anteroposterior displacements of the rib cage and abdomen were measured with two pairs of magnetometer coils. VT was calculated from the sum of these three signals, and was simultaneously, measured using a spirometer or flow meter. A total of 1,111 breaths were analyzed for the resting condition, and 1,163 breaths were analyzed for the exercise condition. We found that VTmag wits highly correlated with VTspiro at rest (R-2 = 0.90) and during exercise (R-2 = 0.79) for pooled data. The slope of this relationship approached the line of identity. The mean percentage differences between VTmag and VTspiro were 10.1 +/- 6.6% at rest and 13.5 +/- 8.6% with exercise. By Bland-Altman analysis, the mean differences between VTmag and VTspiro were 38 mL at resistivity changes in posture, and 182 mL during exercise. TImag and TIspiro values and TEmag and TEspiro values also were highly correlated (R-2 = 0.97 and R-2 = 0.95, respectively) for pooled data. Conclusion: A portable magnetometer system can give useful measures of VT, TI, and TE over a wide range of VTS in sitting, standing, and exercising subjects.
引用
收藏
页码:684 / 691
页数:8
相关论文
共 11 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   ACCURACY OF RESPIRATORY INDUCTIVE PLETHYSMOGRAPHY DURING WAKEFULNESS AND SLEEP IN PATIENTS WITH OBSTRUCTIVE SLEEP-APNEA [J].
CANTINEAU, JP ;
ESCOURROU, P ;
SARTENE, R ;
GAULTIER, C ;
GOLDMAN, M .
CHEST, 1992, 102 (04) :1145-1151
[3]   RELIABILITY OF RESPIRATORY INDUCTIVE PLETHYSMOGRAPHY FOR MEASURING TIDAL VOLUME DURING EXERCISE [J].
CARETTI, DM ;
PULLEN, PV ;
PREMO, LA ;
KUHLMANN, WD .
AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1994, 55 (10) :918-923
[4]   MEASUREMENT OF SEPARATE VOLUME CHANGES OF RIB CAGE AND ABDOMEN DURING BREATHING [J].
KONNO, K ;
MEAD, J .
JOURNAL OF APPLIED PHYSIOLOGY, 1967, 22 (03) :407-&
[5]   ESTIMATES OF VENTILATION FROM BODY-SURFACE MEASUREMENTS IN UNRESTRAINED SUBJECTS [J].
MCCOOL, FD ;
KELLY, KB ;
LORING, SH ;
GREAVES, IA ;
MEAD, J .
JOURNAL OF APPLIED PHYSIOLOGY, 1986, 61 (03) :1114-1119
[6]   THE ROLE OF THE INFERIOR COLLICULAR CORTEX IN THE NEONATAL RAT - SENSORIMOTOR MODULATION [J].
MCCOWN, TJ ;
BREESE, GR .
DEVELOPMENTAL BRAIN RESEARCH, 1991, 59 (01) :1-5
[7]   BREATHING PATTERNS DURING VARIED ACTIVITIES [J].
PAEK, D ;
MCCOOL, FD .
JOURNAL OF APPLIED PHYSIOLOGY, 1992, 73 (03) :887-893
[8]   POSTURAL EFFECTS ON MEASUREMENTS OF TIDAL VOLUME FROM BODY-SURFACE DISPLACEMENTS [J].
PAEK, D ;
KELLY, KB ;
MCCOOL, FD .
JOURNAL OF APPLIED PHYSIOLOGY, 1990, 68 (06) :2482-2487
[9]   3 DEGREE OF FREEDOM DESCRIPTION OF MOVEMENT OF THE HUMAN CHEST-WALL [J].
SMITH, JC ;
MEAD, J .
JOURNAL OF APPLIED PHYSIOLOGY, 1986, 60 (03) :928-934
[10]   ACCURACY OF RESPIRATORY INDUCTIVE PLETHYSMOGRAPH IN MEASURING TIDAL VOLUME DURING SLEEP [J].
WHYTE, KF ;
GUGGER, M ;
GOULD, GA ;
MOLLOY, J ;
WRAITH, PK ;
DOUGLAS, NJ .
JOURNAL OF APPLIED PHYSIOLOGY, 1991, 71 (05) :1866-1871