Optical and atomic force microscopy of an explanted AcrySof intraocular lens with glistenings

被引:52
作者
Dogru, M
Tetsumoto, K
Tagami, Y
Kato, K
Nakamae, K
机构
[1] Shakai Hoken Kobe Cent Hosp, Dept Ophthalmol, Kita Ku, Kobe, Hyogo 6511145, Japan
[2] Kobe Kaisei Hosp, Dept Ophthalmol, Kobe, Hyogo, Japan
[3] Tagami Eye Clin, Kobe, Hyogo, Japan
[4] Kobe Univ, Fac Engn, Kobe, Hyogo, Japan
关键词
D O I
10.1016/S0886-3350(99)00342-9
中图分类号
R77 [眼科学];
学科分类号
100212 [眼科学];
摘要
Purpose: To assess the surface morphology and cause of glistenings in an explanted AcrySof(R) intraocular lens (IOL). Setting: Shakai Hoken Kobe Central Hospital, Kobe, Japan. Methods: A 63-year-old Japanese man had implantation of an AcrySof IOL in the capsular bag. One month postoperatively, he had a nepdymium:YAG laser capsulotomy for posterior capsule opacification, which changed the IOL's position in the capsular bag. A few months later, the patient developed disabling night glare from intralenticular glistenings and progressive hyperopic refractive error. The IOL was explanted and then analyzed by optical microscopy and atomic force microscopy (AFM). Laboratory analysis of control AcrySof IOLs kept in a balanced salt solution at steady room and body temperature for 2 months was also performed to evaluate the cause of the glistenings observed clinically. Results: Optical microscopy showed that the explanted AcrySof IOL had several micro; vacuoles; no abnormalities were observed in the control AcrySof IOLs before or after folding at the room and body temperatures. The AFM analysis showed a significant change in the surface morphology of the explanted IOL, including vacuolar formations in the posterior surface as well as numerous anterior surface irregularities, No microvacuoles or surface morphology alterations were observed in the control AcrySof IOLs by AFM analysis. Conclusion: The glistenings in the explanted AcrySof IOL were likely caused by temperature changes and not mechanical stress from folding. J Cataract Refract Surg 2000; 26:571-575 (C) 2000 ASCRS and ESCRS.
引用
收藏
页码:571 / 575
页数:5
相关论文
共 13 条
[1]
Anderson C, 1993, FOLDABLE INTRAOCULAR, P161
[2]
IMAGING SURFACES OF HYDROPHILIC CONTACT-LENSES WITH THE ATOMIC FORCE MICROSCOPE [J].
BAGUET, J ;
SOMMER, F ;
DUC, TM .
BIOMATERIALS, 1993, 14 (04) :279-284
[3]
Bhatia Sameer, 1997, CLAO Journal, V23, P264
[4]
Visual significance of glistenings seen in the AcrySof intraocular lens [J].
Dhaliwal, DK ;
Mamalis, N ;
Olson, RJ ;
Crandall, AS ;
Zimmerman, P ;
Alldredge, OC ;
Durcan, FJ ;
Omar, O .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1996, 22 (04) :452-457
[5]
Biocompatibility of poly(methyl methacrylate), silicone, and AcrySof intraocular lenses: Randomized comparison of the cellular reaction on the anterior lens surface [J].
Hollick, EJ ;
Spalton, DJ ;
Ursell, PG ;
Pande, MV .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (03) :361-366
[6]
Human cornea and sclera studied by atomic force microscopy [J].
Meller, D ;
Peters, K ;
Meller, K .
CELL AND TISSUE RESEARCH, 1997, 288 (01) :111-118
[7]
Miyata Akira, 1997, Rinsho Ganka, V51, P729
[8]
MORI S, 1998, ATARASHII GANKA, V15, P1011
[9]
Adhesiveness of AcrySof to a collagen film [J].
Nagata, T ;
Minakata, A ;
Watanabe, I .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (03) :367-370
[10]
In vitro analysis of AcrySof intraocular lens glistenings in AcryPak and Wagon Wheel packaging [J].
Omar, O ;
Pirayesh, A ;
Mamalis, N ;
Olson, RJ .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (01) :107-113