The ultrasonic coagulating and cutting system injures nerve function

被引:83
作者
Owaki, T
Nakano, S
Arimura, K
Aikou, T
机构
[1] Kagoshima Univ, Sch Med, Dept Surg 1, Kagoshima 8908520, Japan
[2] Kagoshima Univ, Sch Med, Dept Internal Med 3, Kagoshima 8908520, Japan
关键词
D O I
10.1055/s-2002-33221
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Background: Although, many surgeons have used ultrasonic coagulating and cutting systems and shears (LCS) when carrying out endoscopic thyroidectomies and parathyroidectomies, in our experience some patients had temporary paralysis of the recurrent laryngeal nerve (RLN) after these operations. We had sometimes noticed that the heat of the blade of the LCS was increased just after use; therefore, we designed an RLN model, and investigated the damage to the RLN which could be attributed to the heat of the LCS. Methods: We investigated the effects of the heat of the LCS blade on rat femoral and sciatic nerves, using temperature measurement, histological examination and evoked electromyography. Results: The temperature of the LCS blade exceeded 150degreesC after 30 seconds when it was not used for cutting. When we used the LCS to cut rat muscle or fat tissue, the temperature of the blade exceeded 100degreesC after 20 seconds. There was no damage to the nerve histologically when the LCS was used for less than 20 seconds at a distance of 3 mm. Electrophysiological study showed that touching the nerve with the LCS blade after only 5 seconds of use resulted in damage to the nerve. Conclusions: Our results suggest that the RLN should not be touched directly with the blade just after it has been used, and that it is possible to use the LCS at a distance of 3 mm from the RLN for less than 20 seconds at level 3. In order to maintain these distances, the RLN must be endoscopically visualized during surgery of the neck.
引用
收藏
页码:575 / 579
页数:5
相关论文
共 13 条
[1]  
AMARAL JF, 1995, SURG LAPAROSC ENDOSC, V5, P255
[2]  
AMERAL JF, 1994, SURG LAPARO ENDO PER, V4, P92
[3]   Experimental development of an endoscopic approach to neck exploration and parathyroidectomy [J].
Brunt, LM ;
Jones, DB ;
Wu, JS ;
Quasebarth, MA ;
Meininger, T ;
Soper, NJ .
SURGERY, 1997, 122 (05) :893-901
[5]   Laparoscopic nephrectomy using the Harmonic Scalpel [J].
Helal, M ;
Albertini, J ;
Lockhart, J ;
Albrink, M .
JOURNAL OF ENDOUROLOGY, 1997, 11 (04) :267-268
[6]   Minimally invasive videoscopic parathyroidectomy by lateral approach [J].
Henry, JF ;
Defechereux, T ;
Gramatica, L ;
de Boissezon, C .
LANGENBECKS ARCHIVES OF SURGERY, 1999, 384 (03) :298-301
[7]   Videoendoscopic thyroidectomy: Experimental development of a new technique [J].
Jones, DB ;
Quasebarth, MA ;
Brunt, LM .
SURGICAL LAPAROSCOPY ENDOSCOPY & PERCUTANEOUS TECHNIQUES, 1999, 9 (03) :167-170
[8]  
KANAUCHI H, 1996, ENDOCRINOL J, V45, P135
[9]   Endoscopic endocrine surgery in the neck - An initial report of endoscopic subtotal parathyroidectomy [J].
Naitoh, T ;
Gagner, M ;
Garcia-Ruiz, A ;
Heniford, BT .
SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES, 1998, 12 (03) :202-205
[10]  
Rothenberg SS, 1996, J LAPAROENDOSC SURG, V6, pS61