Biomechanical analysis and quantitative analysis of bone scintigraphy on thrust plate hip prosthesis

被引:9
作者
Goto, T
Yasunaga, Y
Takahashi, K
Ochi, M
机构
[1] Hiroshima Prefectural Akitsu Hosp, Dept Orthopaed Surg, Hiroshima 7292402, Japan
[2] Hiroshima Univ, Sch Med, Dept Orthopaed Surg, Hiroshima, Japan
关键词
strain; total hip arthroplasty; femoral prosthesis; biomechanics; bone scintigraphy;
D O I
10.1007/s00402-0650-0
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction. To study the load distribution of the thrust plate hip prosthesis (TPP, Sulzer Medica, Baar, Switzerland) on femoral bone, a TPP insertion group and intramedullary cemented stem insertion group were prepared with the use of embalmed femoral anatomic specimens, and strain on the femoral bone induced by loading was measured. Materials and methods. Bone scintigraphy was performed in 26 joints which underwent total hip arthroplasty with the use of TPP, to enable quantitative evaluation of postoperative remodeling of the bone. Results. In the TPP insertion group, the strain was almost equivalent to that in the untreated femoral control group. On the other hand, compared with the control group and TPP insertion group, strain in the proximal portion of the femoral bone (medial side of the femoral neck) was significantly smaller in the cemented stem group (control group versus cemented stem insertion group, p=0.0062; cemented stem insertion group versus TPP 130degrees insertion group, p=0.0005; cemented stem insertion group versus TPP 140degrees insertion group, p=0.0031). Bone scintigrams revealed decreased uptake at 12 weeks after operation compared with 6 weeks after operation, indicating that TPP allows earlier implant stabilization than the conventional stem. Conclusion. Compared with the conventional stem, TPP is a prosthesis that can perform physiological load transmission.
引用
收藏
页码:357 / 362
页数:6
相关论文
共 21 条
[1]
BEREITER H, 1991, ORTHOP REL SCI, V2, P191
[2]
Bereiter H, 1997, THRUST PLATE HIP PRO, P48
[3]
BOBYN JD, 1992, CLIN ORTHOP RELAT R, P79
[4]
CANIGGIA M, 1991, Italian Journal of Orthopaedics and Traumatology, V17, P375
[5]
DAntonio JA, 1997, CLIN ORTHOP RELAT R, P124
[6]
Diegel P D, 1989, J Arthroplasty, V4, P173, DOI 10.1016/S0883-5403(89)80071-3
[7]
CALCAR UNLOADING AFTER HIP-REPLACEMENT - A CADAVER STUDY OF FEMORAL STEM DESIGNS [J].
DJERF, K ;
GILLQUIST, J .
ACTA ORTHOPAEDICA SCANDINAVICA, 1987, 58 (02) :97-103
[8]
Dorr L D, 1994, J Arthroplasty, V9, P453, DOI 10.1016/0883-5403(94)90090-6
[9]
POROUS-COATED HIP-REPLACEMENT - THE FACTORS GOVERNING BONE INGROWTH, STRESS SHIELDING, AND CLINICAL-RESULTS [J].
ENGH, CA ;
BOBYN, JD ;
GLASSMAN, AH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1987, 69 (01) :45-55
[10]
Short- and medium-term results of the thrust plate prosthesis in patients with polyarthritis [J].
Fink, B ;
Siegmüller, C ;
Schneider, T ;
Conrad, S ;
Schmielau, G ;
Rüther, W .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2000, 120 (5-6) :294-298