Parameter estimation in dynamic plane source method

被引:37
作者
Malinaric, S [1 ]
机构
[1] Constantine Philosopher Univ, Dept Phys, SK-94974 Nitra, Slovakia
关键词
transient method; thermal diffusivity; thermal conductivity; sensitivity coefficient; difference analysis;
D O I
10.1088/0957-0233/15/5/005
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The paper reports on the measurement of thermophysical properties (thermal conductivity and diffusivity) of PMMA (polymethylmethacrylate). The theory of the dynamic plane heat source method and the experimental apparatus are discussed. Both the sensitivity coefficients and difference analysis are used for searching the time interval in which the measured data can be evaluated. Besides, the influence of the heater and heat losses from the lateral sides of the sample and also the influence of the noise are studied. The estimated measurement uncertainties are 3.6% for thermal conductivity and 2.7% for thermal diffusivity.
引用
收藏
页码:807 / 813
页数:7
相关论文
共 14 条
[1]  
BEC JV, 1977, PARAMETER ESTIMATION
[2]   Parameter estimations for measurements of thermal transport properties with the hot disk thermal constants analyzer [J].
Bohac, V ;
Gustavsson, MK ;
Kubicar, L ;
Gustafsson, SE .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2000, 71 (06) :2452-2455
[3]  
CARSLAW HS, 1959, CONDUCTON HEAT SOLID
[4]   Transient hot strip (THS) method: Uncertainty assessment [J].
Hammerschmidt, U ;
Sabuga, W .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2000, 21 (01) :217-248
[5]   Transient hot wire (THW) method: Uncertainty assessment [J].
Hammerschmidt, U ;
Sabuga, W .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2000, 21 (06) :1255-1278
[6]  
International Organization for Standardization, 1993, GUID EXPR UNC MEAS
[7]   AN EXTENSION TO THE DYNAMIC PLANE SOURCE TECHNIQUE FOR MEASURING THERMAL-CONDUCTIVITY, THERMAL-DIFFUSIVITY, AND SPECIFIC-HEAT OF DIELECTRIC SOLIDS [J].
KARAWACKI, E ;
SULEIMAN, BM ;
ULHAQ, I ;
NHI, BT .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (10) :4390-4397
[8]   DYNAMIC PLANE SOURCE TECHNIQUE FOR SIMULTANEOUS DETERMINATION OF SPECIFIC-HEAT, THERMAL-CONDUCTIVITY AND THERMAL-DIFFUSIVITY OF METALLIC SAMPLES [J].
KARAWACKI, E ;
SULEIMAN, BM .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1991, 2 (08) :744-750
[9]  
Kubicar L., 1999, P 24 INT C THERMAL C, P135
[10]  
MALINARIC S, 2002, THERMOPHYSICS 2002