Measurement of correlated noise in images of computed radiography systems and its influence on the detective quantum efficiency

被引:10
作者
Illers, H [1 ]
Vandenbroucke, D [1 ]
Buhr, E [1 ]
机构
[1] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
来源
MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2 | 2004年 / 5368卷
关键词
computed radiography CR; noise power spectrum NPS; correlated noise; fixed pattern noise; flat field correction; detective quantum efficiency DQE;
D O I
10.1117/12.535807
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
The contributors to image noise of two computed radiography (CR) detector systems - a state-of-the-art and a well-chosen laboratory CR image plate - were studied by two different methods. Method 1 analyzes the image noise content of a series of images obtained at a wide range of different X-ray exposure levels. It uses a model to fit the observed exposure dependence of the normalized noise power spectrum (NNPS): It distinguishes between an NNPS component that is independent of the exposure level and mainly due to correlated noise, and an NNPS component which is inversely proportional to the exposure level and consists mainly of quantum noise. Method 2 analyzes several images taken at the same exposure level and distinguishes between correlated noise, which remains unchanged in repeated exposures, and uncorrelated noise which is different in each image. The results of the two methods allowed the relevant noise contributions in CR images to be quantitatively determined. The novel laboratory image plate showed a significant reduction of correlated noise with an accompanying increase in the DQE. The results also served to estimate a possible improvement of DQE if an appropriate flat field correction is made for these CR systems.
引用
收藏
页码:639 / 647
页数:9
相关论文
共 9 条
[1]
[Anonymous], 622201 IEC 1
[2]
[Anonymous], 1994, 61267 IEC
[3]
Cunningham I. A., 2000, HDB MED IMAGING, V1, P79
[4]
Dobbins JT, 2000, HDB MED IMAGING, P161, DOI DOI 10.1117/3.832716.CH3
[5]
IMAGING PERFORMANCE OF A DIGITAL STORAGE PHOSPHOR SYSTEM [J].
HILLEN, W ;
SCHIEBEL, U ;
ZAENGEL, T .
MEDICAL PHYSICS, 1987, 14 (05) :744-751
[6]
*ISO, 2002, 15739 ISO
[7]
Lubinsky A. R., 1987, Proceedings of the SPIE - The International Society for Optical Engineering, V767, P167
[8]
Rowlands J.A., 2000, Handbook of Medical Imaging, V1, P223, DOI [https://doi.org/10.1117/3.832716, DOI 10.1117/3.832716.CH4, DOI 10.1117/3.832716]
[9]
The physics of computed radiography [J].
Rowlands, JA .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (23) :R123-R166