Modular hip stems: Determination of disassembly force of a neck-stem coupling

被引:15
作者
Pallini, Francesco
Cristofolini, Luca
Traina, Francesco
Toni, Aldo
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, I-40136 Bologna, Italy
[2] Univ Bologna, Dipartimento Ingn Costruz Meccaniche Nucl Aeronau, Bologna, Italy
关键词
hip prosthesis; neck-stem modularity; ceramic coupling; surgical technique; disassembly force;
D O I
10.1111/j.1525-1594.2007.00359.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
An increasing number of hip prostheses feature double head-neck and neck-stem modularities. One possible complication is loosening of the neck-stem coupling. Current surgical technique includes application of hammer blows to the proximal end of the neck-stem coupling. This could compromise the cleanness of the head-neck modularity and damage the bearing surfaces. The goal of this work was to assess whether such hammer blows are really necessary to avoid neck-stem disassembly during in vivo service. Commercially available hip stems featuring neck-stem modularity were tested. Extraction force was measured in vitro when different levels of press fitting were simulated (manual assembly followed by various combinations of simulated loads) to assess critical conditions in relation to surgical technique and applied loads. The disassembly force after manual insertion followed by the first small postoperative loads imposed by the patient during walking was as high as that obtained with hammer blows. Thus, application of hammer blows to fix the neck-stem coupling seemed unnecessary.
引用
收藏
页码:166 / 170
页数:6
相关论文
共 9 条
[1]
*ASTM, 2000, F200900 ASTM INT
[2]
Concerns about ceramics in THA [J].
Barrack, RL ;
Barak, C ;
Skinner, HB .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (429) :73-79
[3]
BERGMANN G, 2001, HIP 98 LOADING HIP J
[4]
Stem modularity is unnecessary in revision total hip arthroplasty - In opposition [J].
Cameron, HU .
JOURNAL OF ARTHROPLASTY, 2003, 18 (03) :101-103
[5]
CONTACT PRESSURES FROM AN INSTRUMENTED HIP ENDOPROSTHESIS [J].
HODGE, WA ;
CARLSON, KL ;
FIJAN, RS ;
BURGESS, RG ;
RILEY, PO ;
HARRIS, WH ;
MANN, RW .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1989, 71A (09) :1378-1386
[6]
Jones RE, 2004, CLIN ORTHOP RELAT R, P142
[7]
The three-dimensional shape of the dysplastic femur [J].
Noble, PC ;
Kamaric, E ;
Sugano, N ;
Matsubara, M ;
Harada, Y ;
Ohzono, K ;
Paravic, V .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (417) :27-40
[8]
Registration of hip prostheses at the Rizzoli institute - 11 years' experience [J].
Stea, S ;
Bordini, B ;
Sudanese, A ;
Toni, A .
ACTA ORTHOPAEDICA SCANDINAVICA, 2002, 73 :40-44
[9]
Effect of femoral head surface roughness on the wear of ultrahigh molecular weight polyethylene acetabular cups [J].
Wang, A ;
Polineni, VK ;
Stark, C ;
Dumbleton, JH .
JOURNAL OF ARTHROPLASTY, 1998, 13 (06) :615-620