Fast position and time sensitive read-out of image intensifiers for single photon detection

被引:16
作者
Jagutzki, O [1 ]
Barnstedt, J [1 ]
Spillmann, U [1 ]
Spielberger, L [1 ]
Mergel, V [1 ]
Ullmann-Pfleger, K [1 ]
Grewing, M [1 ]
Schmidt-Böcking, H [1 ]
机构
[1] Univ Frankfurt, Inst Kernphys, D-60486 Frankfurt, Germany
来源
ULTRAVIOLET AND X-RAY DETECTION, SPECTROSCOPY, AND POLARIMETRY III | 1999年 / 3764卷
关键词
simultaneous position and time-of-flight spectroscopy; microchannel plate; image intensifier; single photon imaging; delay-line technique; UV-detector; fast timing read-out; reconnaissance;
D O I
10.1117/12.371099
中图分类号
V [航空、航天];
学科分类号
08 ; 0825 ;
摘要
We present results on novel image intensifier tubes for single photon detection. We have adopted an image charge coupling technique that allows a read-out of image intensifiers with good imaging properties and much superior time resolution than obtainable with the standard phosphor screen read-out. Although combinations of sealed microchannel plate detector tubes with position and time sensitive anode structures have already been reported, our method has the advantage that the superficial electrode array has not to be implemented inside the tube. We couple the image charge from a high-resistive anode layer through the vacuum housing to a wedge-and-strip or delay-line pattern that can be attached from outside. We show results on single photon imaging with special intensifiers produced by Proxitronic GmbH in the visible and UV for an active diameter of 25 mm. The variability of the system, especially a version with a solar-blind W-cathode and 40 mm active diameter, should open great opportunities for detection tasks in various fields like astronomy, reconnaissance, bioluminescence, atomic physics, and material research, particularly when both good imaging and timing performance are required.
引用
收藏
页码:61 / 69
页数:9
相关论文
共 22 条
[1]   A new type of multiparticle three-dimensional imaging detector with subnanosecond time resolution [J].
Amitay, Z ;
Zajfman, D .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1997, 68 (03) :1387-1392
[2]  
BARNSTEDT J, IN PRESS NUCL INSTRU
[3]   RESISTIVE CATHODE TRANSPARENCY [J].
BATTISTONI, G ;
CAMPANA, P ;
CHIARELLA, V ;
DENNI, U ;
IAROCCI, E ;
NICOLETTI, G .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1982, 202 (03) :459-464
[4]   Performance of an energy-compensated three-dimensional atom probe [J].
Cerezo, A ;
Godfrey, TJ ;
Sijbrandij, SJ ;
Smith, GDW ;
Warren, PJ .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1998, 69 (01) :49-58
[5]   SIMPLE 2-DIMENSIONAL POSITION-SENSITIVE DETECTOR WITH SHORT DEAD-TIME FOR COINCIDENCE EXPERIMENTS [J].
ELAND, JHD .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1994, 5 (12) :1501-1504
[7]   Multilayer anode with crossed serpentine delay lines for high spatial resolution readout of microchannel plate detectors [J].
Friedman, PG ;
Cuza, RA ;
Fleischman, JR ;
Martin, C ;
Schiminovich, D ;
Doyle, DJ .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1996, 67 (02) :596-608
[8]   Fast position and time resolved read-out of micro-channelplates with the delay-line technique for single particle and photon detection [J].
Jagutzki, O ;
Mergel, V ;
Ullmann-Pfleger, K ;
Spielberger, L ;
Meyer, U ;
Dorner, R ;
Schmidt-Bocking, H .
IMAGING SPECTROMETRY IV, 1998, 3438 :322-333
[9]  
JAGUTZKI O, IN PRESS NUCL INSTRU
[10]   A POSITION-SENSITIVE MICROCHANNELPLATE DETECTOR USING A DELAY-LINE READOUT ANODE [J].
KELLER, H ;
KLINGELHOFER, G ;
KANKELEIT, E .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1987, 258 (02) :221-224