MR coil design for simultaneous tip tracking and curvature delineation of a catheter

被引:47
作者
Zuehlsdorff, S
Umathum, R
Volz, S
Hallscheidt, P
Fink, C
Semmler, W
Bock, M
机构
[1] Deutsch Krebsforschungszentrum, D-69120 Heidelberg, Germany
[2] Heidelberg Univ, Dept Radiol, Heidelberg, Germany
关键词
interventional MRI; interactive real-time MRI; intravascular catheters; active device tracking; radio-frequency coil design;
D O I
10.1002/mrm.20108
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In active catheter tracking, small RF coils are attached to the catheter for localization. For interactive catheter steering at vessel branchings, it is necessary to visualize not only a single point near the catheter tip but also the entire shape and orientation of the catheter's distal end. Therefore, a 35-mm-long twisted-pair RF coil was added to a 5 French intravascular catheter with a single tip-tracking coil. With the use of small nonmagnetic electronic components at the catheter tip, and a special switching circuitry outside the catheter, the coil assembly could be operated in two different modes. During MRI, the tip-tracking coil was detuned so that the MR signal was received by the visualization coil only. During tracking, detuning was switched off and the MR signal was predominantly received by the more sensitive tracking coil. The catheter was used in combination with a MR pulse sequence with automatic slice positioning so that the current imaging slice was always placed at the position of the catheter tip. Phantom and animal experiments showed that the catheter tip is better visualized with the combined approach than with a tracking coil alone.
引用
收藏
页码:214 / 218
页数:5
相关论文
共 23 条
[1]  
BOCK M, 2003, P 11 ANN M ISMRM TOR, P231
[2]  
Burl M, 1999, MAGNET RESON MED, V41, P636, DOI 10.1002/(SICI)1522-2594(199903)41:3<636::AID-MRM30>3.0.CO
[3]  
2-A
[4]  
DEBATIN JF, 1999, INTERVENTIONAL MAGNE
[5]   REAL-TIME POSITION MONITORING OF INVASIVE DEVICES USING MAGNETIC-RESONANCE [J].
DUMOULIN, CL ;
SOUZA, SP ;
DARROW, RD .
MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (03) :411-415
[6]   Time-resolved projection angiography after bolus injection of contrast agent [J].
Hennig, J ;
Scheffler, K ;
Laubenberger, J ;
Strecker, R .
MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (03) :341-345
[7]   MR imaging-guided stent placement in iliac arterial stenoses:: A feasibility study [J].
Manke, C ;
Nitz, WR ;
Djavidani, B ;
Strotzer, M ;
Lenhart, M ;
Völk, M ;
Feuerbach, S ;
Link, J .
RADIOLOGY, 2001, 219 (02) :527-534
[8]  
MELZER A, 2000, RADIOLOGY, V217, P913
[9]  
Nitz WR, 2001, J MAGN RESON IMAGING, V13, P105, DOI 10.1002/1522-2586(200101)13:1<105::AID-JMRI1016>3.0.CO
[10]  
2-0