High-performance liquid chromatographic method for the simultaneous determination of the six HIV-protease inhibitors and two non-nucleoside reverse transcriptase inhibitors in human plasma

被引:47
作者
Titier, K
Lagrange, F
Péhourcq, F
Edno-Mcheik, L
Moore, N
Molimard, M
机构
[1] Hop Pellegrin, Dept Clin Pharmacol & Toxicol, Lab Pharmacol Clin & Toxicol, F-33076 Bordeaux, France
[2] Univ Victor Segalen, Bordeaux, France
关键词
liquid chromatography; protease inhibitors; non-nucleoside reverse transcriptase inhibitors;
D O I
10.1097/00007691-200206000-00015
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
A selective and sensitive high-performance liquid chromatographic (HPLC) method has been developed for the determination of the six human immunodeficiency virus (HIV)-protease inhibitors (amprenavir, indinavir, lopinavir, nelfinavir, ritonavir, and saquinavir) and the non-nucleoside reverse transcriptase inhibitors (efavirenz and nevirapine) in a single run. After a liquid-liquid extraction with diethyl ether, the six protease inhibitors and the two non-nucleoside reverse transcriptase inhibitors are separated on a Stability(R) RP18 column eluted with a gradient of acetonitrile and phosphate buffer 50 mmol/L pH 5.65. A sequential ultraviolet detection (5-minute sequence set at 240 nm for nevirapine acquisition, 22-minute sequence set at 215 nm for other antiretroviral drugs acquisition followed by a sequence set at 260 nm for internal standard acquisition) allowed for simultaneous quantitation of the six protease inhibitors, nevirapine, and efavirenz. Calibration curves were linear in the range 100 ng/mL to 10,000 ng/mL. The limit of quantitation was 50 ng/mL for all drugs except nevirapine (100 ng/mL). Average accuracy at four concentrations ranged from 88.2% to 110.9%. Both interday and intraday coefficients of variation were less than 11% for all drugs. The extraction recoveries were greater than 62%. This method is simple and shows a good specificity with respect to commonly co-prescribed drugs, This method allows accurate therapeutic monitoring of amprenavir, indinavir, lopinavir, nelfinavir, ritonavir, saquinavir, efavirenz, and nevirapine.
引用
收藏
页码:417 / 424
页数:8
相关论文
共 27 条
[1]   Indinavir concentrations and antiviral effect [J].
Acosta, EP ;
Henry, K ;
Baken, L ;
Page, LM ;
Fletcher, CV .
PHARMACOTHERAPY, 1999, 19 (06) :708-712
[2]   Determination of twelve antiretroviral agents in human plasma sample using reversed-phase high-performance liquid chromatography [J].
Aymard, G ;
Legrand, M ;
Trichereau, N ;
Diquet, B .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2000, 744 (02) :227-240
[3]   Pharmacokinetics and potential interactions amongst antiretroviral agents used to treat patients with HIV infection [J].
Barry, M ;
Mulcahy, F ;
Merry, C ;
Gibbons, S ;
Back, D .
CLINICAL PHARMACOKINETICS, 1999, 36 (04) :289-304
[4]   Sensitive and rapid method for the simultaneous quantification of the HIV-protease inhibitors indinavir, nelfinavir, ritonavir, and saquinavir in human plasma by reversed-phase liquid chromatography [J].
Bouley, M ;
Briere, C ;
Padoin, C ;
Petitjean, O ;
Tod, M .
THERAPEUTIC DRUG MONITORING, 2001, 23 (01) :56-60
[5]   Rapid quantification of indinavir in human plasma by high-performance liquid chromatography with ultraviolet detection [J].
Foisy, ML ;
Sommadossi, JP .
JOURNAL OF CHROMATOGRAPHY B, 1999, 721 (02) :239-247
[6]   The relationship between ritonavir plasma levels and side-effects: implications for therapeutic drug monitoring [J].
Gatti, G ;
Di Biagio, A ;
Casazza, R ;
De Pascalis, C ;
Bassetti, M ;
Cruciani, M ;
Vella, S ;
Bassetti, D .
AIDS, 1999, 13 (15) :2083-2089
[7]   Treatment with indinavir, zidovudine, and lamivudine in adults with human immunodeficiency virus infection and prior antiretroviral therapy [J].
Gulick, RM ;
Mellors, JW ;
Havlir, D ;
Eron, JJ ;
Gonzalez, C ;
McMahon, D ;
Richman, DD ;
Valentine, FT ;
Jonas, L ;
Meibohm, A ;
Emini, EA ;
Chodakewitz, JA .
NEW ENGLAND JOURNAL OF MEDICINE, 1997, 337 (11) :734-739
[8]   Determination of saquinavir in human plasma, saliva, and cerebrospinal fluid by ion-pair high-performance liquid chromatography with ultraviolet detection [J].
Hoetelmans, RMW ;
vanEssenberg, M ;
Meenhorst, PL ;
Mulder, JW ;
Beijnen, JH .
JOURNAL OF CHROMATOGRAPHY B, 1997, 698 (1-2) :235-241
[9]   Determination of nevirapine, an HIV-1 non-nucleoside reverse transcriptase inhibitor, in human plasma by reversed-phase high-performance liquid chromatography [J].
Hollanders, RMF ;
Kolmer, EWJV ;
Burger, DM ;
Wuis, EW ;
Koopmans, PP ;
Hekster, YA .
JOURNAL OF CHROMATOGRAPHY B, 2000, 744 (01) :65-71
[10]   Simultaneous determination of the HIV-protease inhibitors indinavir, nelfinavir, saquinavir and ritonavir in human plasma by reversed-phase high-performance liquid chromatography [J].
Hugen, PWH ;
Verweij-van Wissen, CPWGM ;
Burger, DM ;
Wuis, EW ;
Koopmans, PP ;
Hekster, YA .
JOURNAL OF CHROMATOGRAPHY B, 1999, 727 (1-2) :139-149