The absolute signal-to-noise ratio in MRI acceptance testing

被引:7
作者
McRobbie, DW
机构
[1] Radiological Sciences Unit, Hammersmith Hospitals NHS Trust, Charing Cross Hospital, London W6 8RF, Fulham Palace Road
关键词
D O I
10.1259/0007-1285-69-827-1045
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The absolute signal-to-noise ratio (SNR) measurable from images is defined and related to the theoretical fundamental SNR for MRI. Absolute SNR for quadrature head coils was obtained by both measurement and analysis of published data for MRI systems in the range 0.5-1.5 T. Mean absolute SNR for 22 MRI systems was 1.59 +/- 0.25 x 10(4) Hz(1/2) ml(-1) T-1 and did not vary significantly with the static field strength or gradient strength and signal bandwidth. This can be related to the fundamental SNR calculated as 2.36 x 10(4) Hz(1/2) ml(-1) T-1 through an effective noise figure with an average value of 3.4 dB for the 22 systems. Use of this methodology is discussed in relation to acceptance testing and performance assessment.
引用
收藏
页码:1045 / 1048
页数:4
相关论文
共 27 条
[1]   A SIGNAL-TO-NOISE CALIBRATION PROCEDURE FOR NMR IMAGING-SYSTEMS [J].
EDELSTEIN, WA ;
BOTTOMLEY, PA ;
PFEIFER, LM .
MEDICAL PHYSICS, 1984, 11 (02) :180-185
[2]   THE INTRINSIC SIGNAL-TO-NOISE RATIO IN NMR IMAGING [J].
EDELSTEIN, WA ;
GLOVER, GH ;
HARDY, CJ ;
REDINGTON, RW .
MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (04) :604-618
[3]   SENSITIVITY OF THE ZEUGMATOGRAPHIC EXPERIMENT INVOLVING HUMAN SAMPLES [J].
HOULT, DI ;
LAUTERBUR, PC .
JOURNAL OF MAGNETIC RESONANCE, 1979, 34 (02) :425-433
[4]   SIMULATION OF PATIENT LOADING IN NUCLEAR-MAGNETIC-RESONANCE IMAGING THROUGH THE USE OF INDUCTIVE DAMPING LOOPS [J].
JENNINGS, SJ ;
LERSKI, RA ;
MCROBBIE, DW ;
STRAUGHAN, K ;
ORR, JS .
BRITISH JOURNAL OF RADIOLOGY, 1986, 59 (706) :1031-1034
[5]   MULTI-CENTER TRIAL WITH PROTOCOLS AND PROTOTYPE TEST OBJECTS FOR THE ASSESSMENT OF MRI EQUIPMENT [J].
LERSKI, RA ;
MCROBBIE, DW ;
STRAUGHAN, K ;
WALKER, PM ;
DECERTAINES, JD ;
BERNARD, AM .
MAGNETIC RESONANCE IMAGING, 1988, 6 (02) :201-214
[6]   RADIO-FREQUENCY EDDY-CURRENT LOSSES FOR AN ANNULAR CONDUCTOR IN MRI - THEORY AND APPLICATIONS [J].
MCROBBIE, DW .
MEDICAL PHYSICS, 1993, 20 (05) :1555-1560
[7]  
*MDA, 1994, MDA9473
[8]  
*MDA, 1994, MDA9472
[9]  
*MDA, 1994, MDA9471
[10]  
*MDA, 1994, MDA9470