Are neutrons responsible for the dose discrepancies between Monte Carlo calculations and measurements in the build-up region for a high-energy photon beam?

被引:11
作者
Ding, GX [1 ]
Duzenli, C [1 ]
Kalach, NI [1 ]
机构
[1] British Columbia Canc Agcy, Faser Valley Canc Ctr, Surrey, BC V3V 1Z2, Canada
关键词
D O I
10.1088/0031-9155/47/17/313
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study presents measured neutron dose using a neutron dosimeter in a water phantom and investigates a hypothesis that neutrons in a high-energy photon beam may be responsible for the reported significant dose discrepancies between Monte Carlo calculations and measurements at the build-up region in large fields. Borated polyethylene slabs were inserted between the accelerator head and the phantom in order to remove neutrons generated in the accelerator head. The thickness of the slab ranged from 2.5 cm to 10 cm. A lead slab of 3 mm thickness was also used in the study. The superheated drop neutron dosimeter was used to measure the depth-dose curve of neutrons in a. high-energy photon beam and to verify the effectiveness of the slab to remove these neutrons. Total dose measurements were performed in water using a WELLHOFER WP700 beam scanner with an IC-10 ionization chamber. The Monte Carlo code BEAM was used to simulate an 18 MV photon beam from a Varian Clinac-2100EX accelerator. Both EGS4/DOSXYZ and EGSnrc/DOSRZnrc were used in the dose calculations. Measured neutron dose equivalents as a function of depth per unit total dose in water were presented for 10 x 10 and 40 x 40 cm(2) fields. The measured results have shown that a 5-10 cm thick borated polyethylene slab can reduce the neutron dose by a factor of 2 when inserted between the accelerator head and the detector. In all cases the measured neutron dose equivalent was less than 0.5% of the photon dose. In order to study if the ion chamber was highly sensitive to the neutron dose, we have investigated the disagreement between the Monte Carlo calculated and measured central-axis depth-dose curves in the build-up region when different shielding materials were used. The result indicated that the IC-10 chamber was not highly sensitive to the neutron dose. Therefore, neutrons present in a high-energy photon beam were unlikely to be responsible for the reported discrepancies in the build-up region for large fields.
引用
收藏
页码:3251 / 3261
页数:11
相关论文
共 19 条
[1]   AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams [J].
Almond, PR ;
Biggs, PJ ;
Coursey, BM ;
Hanson, WF ;
Huq, MS ;
Nath, R ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (09) :1847-1870
[2]  
Andreo P, 2000, IAEA TECHNICAL REPOR, V398
[3]  
[Anonymous], 1983, MED PHYS, V10, P741
[4]  
ATTIX F. H., 1986, Introduction to radiological physics and radiation dosimetry
[5]   In-phantom dosimetry and spectrometry of photoneutrons from an 18 MV linear accelerator [J].
d'Errico, F ;
Nath, R ;
Tana, L ;
Curzio, G ;
Alberts, WG .
MEDICAL PHYSICS, 1998, 25 (09) :1717-1724
[6]   Energy spectra, angular spread, fluence profiles and dose distributions of 6 and 18 MV photon beams: results of Monte Carlo simulations for a Varian 2100EX accelerator [J].
Ding, GX .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (07) :1025-1046
[7]   Determination of percentage depth-dose curves for electron beams using different types of detectors [J].
Ding, GX ;
Yu, CW .
MEDICAL PHYSICS, 2001, 28 (03) :298-302
[8]  
DING GX, 1995, THESIS CARLETON U OT
[9]  
DING GX, 2001, INT WORKSH REC DEV A
[10]  
ICRP, 1991, Report No. 60