Realization of matching conditions for high-resolution spectrometers

被引:115
作者
Fujita, H
Fujita, Y
Berg, GPA
Bacher, AD
Foster, CC
Hara, K
Hatanaka, K
Kawabata, T
Noro, T
Sakaguchi, H
Shimbara, Y
Shinada, T
Stephenson, EJ
Ueno, H
Yosoi, M
机构
[1] Osaka Univ, Dept Phys, Toyonaka, Osaka 5600043, Japan
[2] Indiana Univ, IUCF, Bloomington, IN 47408 USA
[3] Osaka Univ, RCNP, Ibaraki, Osaka 5670047, Japan
[4] Kyoto Univ, Dept Phys, Kyoto 6068502, Japan
关键词
magnetic spectrometer; high resolution; large dispersion; kinematic correction; focus matching; lateral dispersion matching; angular dispersion matching;
D O I
10.1016/S0168-9002(01)01970-2
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
For precise measurements of nuclear-reaction spectra using a beam from an accelerator, a high-resolution magnetic spectrometer is a powerful tool. The full capability of a magnetic spectrometer, however, can be achieved only if the characteristics of the beam coming from the accelerator are matched to those required by the spectrometer by properly, adjusting the beam line conditions. The matching methods, lateral dispersion matching, focus matching and also the kinematic correction compensate the spectrum line-broadening effects caused by the beam momentum spread and reaction kinematics. In addition, angular dispersion matching should be performed if good resolution of the scattering angle is important. Diagnostic methods developed to realize these matching conditions for the spectrometers K600 at IUCF and Grand Raiden at RCNP are presented. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:17 / 26
页数:10
相关论文
共 22 条
[1]  
BERG GPA, UNPUB IUCF SCI TECHN, P152
[2]   RECONSTRUCTIVE CORRECTION OF ABERRATIONS IN NUCLEAR-PARTICLE SPECTROGRAPHS [J].
BERZ, M ;
JOH, K ;
NOLEN, JA ;
SHERRILL, BM ;
ZELLER, AF .
PHYSICAL REVIEW C, 1993, 47 (02) :537-544
[3]  
Brown K.L., 1980, CERN8004
[4]   RESOLUTION OF ACCELERATOR MAGNETIC ANALYZING SYSTEMS [J].
COHEN, BL .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1959, 30 (06) :415-418
[5]   BROAD-RANGE PARTICLE SPECTROGRAPHS [J].
ENGE, HA .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1981, 187 (01) :1-11
[6]   MAGNETIC SPECTROGRAPHS FOR NUCLEAR-REACTION STUDIES [J].
ENGE, HA .
NUCLEAR INSTRUMENTS & METHODS, 1979, 162 (1-3) :161-180
[7]  
FOSTER CC, IUCF SCI TECHNOLOGY, P193
[8]   Better-resolution measurement of vertical scattering angle in a new ion-optical mode of spectrometer "Grand Raiden" [J].
Fujita, H ;
Berg, GPA ;
Fujita, Y ;
Hatanaka, K ;
Noro, T ;
Stephenson, EJ ;
Foster, CC ;
Sakaguchi, H ;
Itoh, M ;
Taki, T ;
Tamura, K ;
Ueno, H .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 469 (01) :55-62
[9]  
FUJITA H, RCNP ANN REPORT 1999, P87
[10]   THE QDMDQ MAGNETIC SPECTROGRAPH RAIDEN .2. CORRECTION OF KINEMATIC LINE BROADENING [J].
FUJITA, Y ;
MORINOBU, S ;
FUJIWARA, M ;
KATAYAMA, I ;
YAMAZAKI, T ;
IKEGAMI, H .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1984, 225 (02) :298-307