The press-fit condylar modular total knee system - Four-to-six-year results with a posterior-cruciate-substituting design

被引:82
作者
Ranawat, CS
Luessenhop, CP
Rodriguez, JA
机构
[1] Center for Total Joint Replacement, Lenox Hill Hospital, William Black Hall, New York, NY 10021
[2] Clairemont, OK 74017
[3] Center for Total Joint Replacement, Lenox Hill Hospital, New York City, NY
关键词
D O I
10.2106/00004623-199703000-00004
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
One hundred and eighteen patients who had had 150 consecutive primary total knee replacements (sixteen bilateral procedures) between February 1988 and February 1990, with insertion of the press-fit condylar modular total knee system with cement, were enrolled in a prospective study. Ninety-six patients (125 knees) were followed for an adequate interval (mean, 4.8 years; range, 3.8 to 6.2 years). Thirteen patients (fifteen knees) died, and nine patients (ten knees) were lost to follow-up. The mean age of the patients at the time of the index arthroplasty was seventy years (range, twenty-nine to eighty-five years). The patients were evaluated clinically and radiographically, according to the scoring system of the Knee Society, and the results on a self-administered questionnaire were used to evaluate pain, function, satisfaction, and patellofemoral symptoms. A Kaplan-Meier survivorship analysis was performed with a revision operation as the end point. The mean functional and clinical scores, according to the system of the Knee Society, were 78 and 93 points, respectively, at the most recent follow-up examination. The result was excellent for 103 knees, good for thirteen, fair for three, and poor for six. Three revision operations were necessary: two because of infection and one because of instability. The over-all rate of patellofemoral symptoms was 8 per cent (ten knees). Three knees had tibiofemoral instability; subsequent modification of the design of the tibial cam decreased the prevalence of this problem. Non-progressive radiolucent lines were present at the cement-bone interface in 39 per cent (thirty-nine) of the ninety-nine knees that had complete radiographic follow-up. No prosthesis had loosened by the time of the most recent follow-up examination. The rate of survival of the implant was 97 per cent at six years, and the standard error of the mean was 1.6 per cent. In the present series, total knee arthroplasties with the press-fit condylar modular knee system resulted in excellent relief of pain, an excellent range of motion, and restoration of function. They were also associated with a low prevalence of patellofemoral problems.
引用
收藏
页码:342 / 348
页数:7
相关论文
共 34 条
[1]  
Beight J.L., 1994, CLIN ORTHOP RELAT R, V299, P139
[2]  
BERGER RA, 1995, ORTHOP T, V19, P27
[3]  
BINDELGLASS DF, 1993, CLIN ORTHOP RELAT R, V286, P103
[4]  
Charnley J., 1979, Low friction arthroplasty of the hip: theory and practice, P20
[5]   THE POSTERIOR STABILIZED TOTAL KNEE PROSTHESIS - ASSESSMENT OF POLYETHYLENE DAMAGE AND OSTEOLYSIS AFTER A 10-YEAR-MINIMUM FOLLOW-UP [J].
COLIZZA, WA ;
INSALL, JN ;
SCUDERI, GR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77A (11) :1713-1720
[6]  
DENNIS DA, 1992, CLIN ORTHOP RELAT R, V281, P168
[7]  
EWALD FC, 1989, CLIN ORTHOP RELAT R, P9
[8]   THE INFLUENCE OF TIBIAL-PATELLOFEMORAL LOCATION ON FUNCTION OF THE KNEE IN PATIENTS WITH THE POSTERIOR STABILIZED CONDYLAR KNEE PROSTHESIS [J].
FIGGIE, HE ;
GOLDBERG, VM ;
HEIPLE, KG ;
MOLLER, HS ;
GORDON, NH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1035-1040
[9]  
GRACE JN, 1988, CLIN ORTHOP RELAT R, V237, P184
[10]   TOTAL CONDYLAR KNEE PROSTHESIS - REPORT OF 220 CASES [J].
INSALL, J ;
SCOTT, WN ;
RANAWAT, CS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1979, 61 (02) :173-180