Temperature-controlled radiofrequency catheter ablation with a 10-mm tip electrode creates larger lesions without charring in the porcine heart

被引:9
作者
Anfinsen, OG [1 ]
Aass, H
Kongsgaard, E
Foerster, A
Scott, H
Amlie, JP
机构
[1] Rikshosp, Dept Cardiol, N-0027 Oslo, Norway
[2] Rikshosp, Inst Surg Res, N-0027 Oslo, Norway
[3] Rikshosp, Dept Pathol, N-0027 Oslo, Norway
关键词
catheter ablation; temperature; large electrode; porcine heart; IVC-TV isthmus; atrium;
D O I
10.1023/A:1009840004782
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Radiofrequency catheter ablation of atrial flutter, atrial fibrillation or ventricular tachycardia may be favoured by large lesions. We compared lesions created in unipolar mode using 10-mm/8 F electrodes with those of 4-mm/7 F catheters. Methods: Ablations were first performed in porcine hearts in vitro (70 degrees C, 60 s, tangential catheter tip-tissue orientation). Anaesthetized pigs were thereafter ablated with 10- or 4-mm catheters in the right atrial free wall (RAFW), inferior vena cava-tricuspid valve (IVC-TV) isthmus and left ventricle (LV). Results: In vitro, lesion length doubled and lesion volume tripled using the 10-mm catheter. Average power supply was 69 (SD12) (10-mm tip) versus 26 (SD7) W (4-mm tip). In vivo, lesion length increased by 50% and lesion volume fivefold. Charring at the lesion surface or sudden impedance rises were not observed in vivo. Histologically, coagulation necrosis and minor haemorrhages were found. One RAFW lesion (10-mm) showed a dissection approaching the epicardium. Fibrinous platelet clots or overt thromboses covered the endocardial surface in half of all lesions. Three 10-mm electrode isthmus lesions extended to the right descending posterior artery and one LV lesion to the left anterior descending artery, but there was no damage to the arterial walls. Following six ablations with the 10-mm electrode and two with the 4-mm tip, injury to the adjacent lung tissue of 0.5 to 6.0 mm depth was found (p = 0.22). Conclusion: RF ablation using 10-mm/8 F electrodes created significantly larger lesions. 10-mm electrodes appeared safe in the porcine IVC-TV isthmus and LV, but not in the RAFW.
引用
收藏
页码:343 / 351
页数:9
相关论文
共 36 条
[1]   Radiofrequency current ablation of porcine right atrium: Increased lesion size with bipolar two catheter technique compared to unipolar application in vitro and in vivo [J].
Anfinsen, OG ;
Kongsgaard, E ;
Foerster, A ;
Aass, H ;
Amlie, JP .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (01) :69-78
[2]   Bipolar radiofrequency catheter ablation creates confluent lesions at a larger interelectrode spacing than does unipolar ablation from two electrodes in the porcine heart [J].
Anfinsen, OG ;
Kongsgaard, E ;
Foerster, A ;
Amlie, JP ;
Aass, H .
EUROPEAN HEART JOURNAL, 1998, 19 (07) :1075-1084
[3]  
ANFINSEN OG, 1996, EUR J C P E, V6, P195
[4]   Electrophysiological effects of catheter ablation of inferior vena cava-tricuspid annulus isthmus in common atrial flutter [J].
Cauchemez, B ;
Haissaguerre, M ;
Fischer, B ;
Thomas, O ;
Clementy, J ;
Coumel, P .
CIRCULATION, 1996, 93 (02) :284-294
[5]  
Chen SA, 1998, CIRCULATION, V98, P19
[6]   Radiofrequency catheter ablation of common atrial flutter: Comparison of electrophysiologically guided focal ablation technique and linear ablation technique [J].
Chen, SA ;
Chiang, CE ;
Wu, TJ ;
Tai, CT ;
Lee, SH ;
Cheng, CC ;
Chiou, CW ;
Ueng, KC ;
Wen, ZC ;
Chang, MS .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1996, 27 (04) :860-868
[7]   RADIOFREQUENCY ABLATION OF THE INFERIOR VENA-CAVA - TRICUSPID-VALVE ISTHMUS IN COMMON ATRIAL-FLUTTER [J].
COSIO, FG ;
LOPEZGIL, M ;
GOICOLEA, A ;
ARRIBAS, F ;
BARROSO, JL .
AMERICAN JOURNAL OF CARDIOLOGY, 1993, 71 (08) :705-709
[8]   RADIOFREQUENCY CATHETER ABLATION OF COMMON ATRIAL-FLUTTER IN 80 PATIENTS [J].
FISCHER, B ;
HAISSAGUERRE, M ;
GARRIGUES, S ;
POQUET, F ;
GENCEL, L ;
CLEMENTY, J ;
MARCUS, FI .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1995, 25 (06) :1365-1372
[9]  
Hahn N, 1996, BERL MUNCH TIERARZTL, V109, P23
[10]   ELECTRODE RADIUS PREDICTS LESION RADIUS DURING RADIOFREQUENCY ENERGY HEATING - VALIDATION OF A PROPOSED THERMODYNAMIC MODEL [J].
HAINES, DE ;
WATSON, DD ;
VEROW, AF .
CIRCULATION RESEARCH, 1990, 67 (01) :124-129