Comparison of friction and lubrication of different hip prostheses

被引:116
作者
Scholes, SC [1 ]
Unsworth, A [1 ]
机构
[1] Univ Durham, Sch Engn, Ctr Biomed Engn, Durham DH1 3LE, England
关键词
friction; lubrication; alumina; cobalt-chromium-molybdenum; ultra-high molecular weight polyethylene;
D O I
10.1243/0954411001535237
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
It is well documented that an important cause of osteolysis and subsequent loosening of replacement hip joints is polyethylene wear debris. To avoid this, interest has been renewed in metal-on-metal and ceramic-on-ceramic prostheses. Various workers have assessed the lubrication modes of different joints by measuring the friction at the bearing surfaces, using different lubricants. Measurements of friction factors of a series of hip prostheses were undertaken using carboxymethyl cellulose (CMC) fluids, silicone fluids, synovial fluid and different concentrations of bovine serum as the lubricant. The experimental results were compared with theoretical predictions of film thicknesses and lubrication modes. A strong correlation was observed between experiment and theory when employing CMC fluids or silicone fluids as the lubricant. Mixed lubrication was found to occur in the metal-on-metal (CoCrMo/CoCrMo) joints with all lubricants at a viscosity within the physiological range. This was also the case for the metal-on-plastic (CoCrMo/ultra-high molecular weight polyethylene) joints. The ceramic-on-ceramic (Al2O3/Al2O3) joints, however, exhibited full fluid film lubrication with the synthetic lubricants but mixed lubrication with the biological lubricants. Employing a biological fluid as the lubricant affected the friction to varying degrees when compared with the synthetic lubricants. In the case of the ceramic-on-ceramic joints it acted to increase the friction factor tenfold; however, for the metal-on-metal joints, biological fluids gave slightly lower friction than the synthetic lubricants did. This suggests that, when measuring friction and wear of artificial joints, a standard lubricant should be used.
引用
收藏
页码:49 / 57
页数:9
相关论文
共 40 条
[1]  
CLARKE IC, 1991, TOTAL HIP ARTHROPLAS, P535
[2]  
Cooper J. R., 1993, Clinical Materials, V14, P295, DOI 10.1016/0267-6605(93)90016-Z
[3]  
COOPER JR, 1992, WEAR PARTICLES
[4]  
Fisher J., 1994, CURR ORTHOPAED, V8, P164, DOI DOI 10.1016/0268-0890(94)90048-5
[5]  
GORE TA, 1981, ENG MED, V10, P89
[6]  
Hall RM, 1997, BRIT J RHEUMATOL, V36, P20
[7]   FRICTIONAL CHARACTERIZATION OF EXPLANTED CHARNLEY HIP PROSTHESES [J].
HALL, RM ;
UNSWORTH, A ;
WROBLEWSKI, BM ;
BURGESS, IC .
WEAR, 1994, 175 (1-2) :159-166
[8]   ELASTOHYDRODYNAMIC LUBRICATION OF ELLIPTICAL CONTACTS FOR MATERIALS OF LOW ELASTIC-MODULUS I - FULLY FLOODED CONJUNCTION [J].
HAMROCK, BJ ;
DOWSON, D .
JOURNAL OF LUBRICATION TECHNOLOGY-TRANSACTIONS OF THE ASME, 1978, 100 (02) :236-245
[9]   A RAT MODEL OF RESORPTION OF BONE AT THE CEMENT-BONE INTERFACE IN THE PRESENCE OF POLYETHYLENE WEAR PARTICLES [J].
HOWIE, DW ;
VERNONROBERTS, B ;
OAKESHOTT, R ;
MANTHEY, B .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (02) :257-263
[10]   Analysis of fluid film lubrication in artificial hip joint replacements with surfaces of high elastic modulus [J].
Jin, ZM ;
Dowson, D ;
Fisher, J .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1997, 211 (03) :247-256