Additional luminal area gain by intravascular ultrasound guidance after coronary stent implantation with high inflation pressure

被引:16
作者
Werner, GS
Diedrich, J
Schunemann, S
Gastmann, O
Ferrari, M
Buchwald, AB
Figulla, HR
Kreuzer, H
机构
[1] Georg-August-University,Department of Cardiology, Center for Internal Medicine
来源
INTERNATIONAL JOURNAL OF CARDIAC IMAGING | 1997年 / 13卷 / 04期
关键词
intravascular ultrasound; coronary stent; high-pressure balloon;
D O I
10.1023/A:1005703626872
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims: Studies by intravascular ultrasound demonstrated inadequate expansion in a large number of stents, which lead to the increase of inflation pressures for stenting. The present study examined whether routine use of high-pressure inflation would be sufficient for an optimum stent expansion without sonographic guidance. Methods and results: Two types of single coronary stents (Palmaz-Schatz in 54, and Wiktor in 25) were implanted with inflation pressures of 16-20 atm in 79 nonocclusive coronary lesions. IVUS before stenting was used in 78% to select the adequate stent size. Intravascular ultrasound after stenting was used to assess the minimum stent area and diameter, the reference areas, and the strut apposition to the vessel wall. The difference between the area of the expanding balloon and the stent area was calculated as the luminal deficit of the stent. Completeness of stent expansion required full strut apposition and lesion coverage, and a minimum stent area that was larger than the distal reference, and larger than 60% of the proximal reference. Intravascular ultrasound before stenting lead to an increase of the stent size in 47%. After high-pressure expansion, even with the optimized balloon size, 8% of stents had struts protruding into the lumen. The stent area (6.87 +/- 1.93 mm(2)) was significantly smaller than both the proximal (9.59 +/- 2.91 mm(2); p<0.001) and distal reference area (8.23 +/- 3.03 mm(2); p<0.001). The criteria for complete expansion were met in 48%. The expansion with a larger high-pressure balloon in 25 stents lead to an increase of the stent area by 19% (8.19 +/- 2.24; p<0.001), and full stent apposition in all cases. The criteria of stent expansion were met in 82%. A wide range of the luminal deficit upto 48% was observed, which was not related to sonographic lesion characteristics, except in lesions with complete circumferential calcifications. The different stent designs were characterized by a slightly lower luminal deficit in slotted-tube stents (23 +/- 13% vs. 28 +/- 12%; p = 0.11) and a better index of stent symmetry as compared with the coil stent (0.87 +/- 0.08 vs. 0.82 +/- 0.09; p<0.05). Conclusion: Routine use of high-pressure stent expansion did not lead to a sufficient stent expansion, even when the initial stent size had been guided by intravascular ultrasound. Further stent dilatation with larger balloons under ultrasound guidance would be required to optimize the luminal area gain.
引用
收藏
页码:311 / 321
页数:11
相关论文
共 29 条
[1]  
Buchwald A. B., 1996, European Heart Journal, V17, P172
[2]   INTRACORONARY STENTING WITHOUT ANTICOAGULATION ACCOMPLISHED WITH INTRAVASCULAR ULTRASOUND GUIDANCE [J].
COLOMBO, A ;
HALL, P ;
NAKAMURA, S ;
ALMAGOR, Y ;
MAIELLO, L ;
MARTINI, G ;
GAGLIONE, A ;
GOLDBERG, SL ;
TOBIS, JM .
CIRCULATION, 1995, 91 (06) :1676-1688
[3]  
De Jaegere Peter, 1996, Journal of the American College of Cardiology, V27, p137A
[4]   SMALL STENT SIZE AND INTIMAL HYPERPLASIA CONTRIBUTE TO RESTENOSIS - A VOLUMETRIC INTRAVASCULAR ULTRASOUND ANALYSIS [J].
DUSSAILLANT, GR ;
MINTZ, GS ;
PICHARD, AD ;
KENT, KM ;
SATLER, LF ;
POPMA, JJ ;
WONG, SC ;
LEON, MB .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1995, 26 (03) :720-724
[5]   RESTENOSIS AFTER PLACEMENT OF PALMAZ-SCHATZ STENTS IN NATIVE CORONARY-ARTERIES - INITIAL RESULTS OF A MULTICENTER EXPERIENCE [J].
ELLIS, SG ;
SAVAGE, M ;
FISCHMAN, D ;
BAIM, DS ;
LEON, M ;
GOLDBERG, S ;
HIRSHFELD, JW ;
CLEMAN, MW ;
TEIRSTEIN, PS ;
WALKER, C ;
BAILEY, S ;
BUCHBINDER, M ;
TOPOL, EJ ;
SCHATZ, RA .
CIRCULATION, 1992, 86 (06) :1836-1844
[6]   A RANDOMIZED COMPARISON OF CORONARY-STENT PLACEMENT AND BALLOON ANGIOPLASTY IN THE TREATMENT OF CORONARY-ARTERY DISEASE [J].
FISCHMAN, DL ;
LEON, MB ;
BAIM, DS ;
SCHATZ, RA ;
SAVAGE, MP ;
PENN, I ;
DETRE, K ;
VELTRI, L ;
RICCI, D ;
NOBUYOSHI, M ;
CLEMAN, M ;
HEUSER, R ;
ALMOND, D ;
TEIRSTEIN, PS ;
FISH, RD ;
COLOMBO, A ;
BRINKER, J ;
MOSES, J ;
SHAKNOVICH, A ;
HIRSHFELD, J ;
BAILEY, S ;
ELLIS, S ;
RAKE, R ;
GOLDBERG, S .
NEW ENGLAND JOURNAL OF MEDICINE, 1994, 331 (08) :496-501
[7]   CLASSIFICATION OF MORPHOLOGIC EFFECTS OF PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY ASSESSED BY INTRAVASCULAR ULTRASOUND [J].
GERBER, TC ;
ERBEL, R ;
GORGE, G ;
GE, J ;
RUPPRECHT, HJ ;
MEYER, J .
AMERICAN JOURNAL OF CARDIOLOGY, 1992, 70 (20) :1546-1554
[8]   BENEFIT OF INTRACORONARY ULTRASOUND IN THE DEPLOYMENT OF PALMAZ-SCHATZ STENTS [J].
GOLDBERG, SL ;
COLOMBO, A ;
NAKAMURA, S ;
ALMAGOR, Y ;
MAIELLO, L ;
TOBIS, JM .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 24 (04) :996-1003
[9]   INTRAVASCULAR ULTRASOUND AFTER LOW AND HIGH INFLATION PRESSURE CORONARY-ARTERY STENT IMPLANTATION [J].
GORGE, G ;
HAUDE, M ;
VOEGELE, E ;
GERBER, T ;
RUPPRECHT, HJ ;
MEYER, J ;
ERBEL, R .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1995, 26 (03) :725-730
[10]   THE SAFETY OF INTRACORONARY ULTRASOUND - A MULTICENTER SURVEY OF 2207 EXAMINATIONS [J].
HAUSMANN, D ;
ERBEL, R ;
ALIBELLICHEMARIN, MJ ;
BOKSCH, W ;
CARACCIOLO, E ;
COHN, JM ;
CULP, SC ;
DANIEL, WG ;
DESCHEERDER, I ;
DIMARIO, C ;
FERGUSON, JJ ;
FITZGERALD, PJ ;
FRIEDRICH, G ;
GE, JB ;
GORGE, G ;
HANRATH, P ;
HODGSON, JM ;
ISNER, JM ;
JAIN, S ;
MAIERRUDOLPH, W ;
MOONEY, M ;
MOSES, JW ;
MUDRA, H ;
PINTO, FJ ;
SMALLING, RW ;
TALLEY, JD ;
TOBIS, JM ;
WALTER, PD ;
WEIDINGER, F ;
WERNER, GS ;
YEUNG, AC ;
YOCK, PG .
CIRCULATION, 1995, 91 (03) :623-630