The relationship of recirculation to access blood flow

被引:96
作者
Besarab, A [1 ]
Sherman, R [1 ]
机构
[1] UNIV MED & DENT NEW JERSEY,ROBERT WOOD JOHNSON MED SCH,DEPT MED,DIV NEPHROL,NEW BRUNSWICK,NJ 08903
关键词
vascular access; recirculation; access blood flow;
D O I
10.1016/S0272-6386(97)90033-0
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
The relationship between vascular access recirculation and access blood flow has been obscured by measurement techniques that overestimate recirculation, We measured brachial artery blood flow (a surrogate for access blood flow) using Doppler ultrasound and access recirculation using a two-needle slow/stop flow method and the standard peripheral vein method in 77 chronic hemodialysis patients, These patients had 25 native arteriovenous fistulae and 52 polytetrafluoroethylene grafts. Access recirculation by the slow/stop flow method was uniformly absent in these patients unless their access blood flow rate was less than the dialyzer blood flow rate, The peripheral vein method averaged 10.7% in these patients; many values were considerably higher, We conclude that access recirculation is a function of access blood flow and does not occur (in the absence of or reversed needle cannulation) unless access blood flow is markedly impaired. (C) 1997 by the National Kidney Foundation, Inc.
引用
收藏
页码:223 / 229
页数:7
相关论文
共 40 条
  • [1] BESARAB A, 1995, J AM SOC NEPHROL, V6, P483
  • [2] Besarab A., 1995, J AM SOC NEPHROL, V7, P483
  • [3] BUUR T, 1994, NEPHROL DIAL TRANSPL, V9, P395
  • [4] FISTULA DYSFUNCTION - EFFECT ON RAPID HEMODIALYSIS
    COLLINS, DM
    LAMBERT, MB
    MIDDLETON, JP
    PROCTOR, RK
    DAVIDSON, CJ
    NEWMAN, GE
    SCHWAB, SJ
    [J]. KIDNEY INTERNATIONAL, 1992, 41 (05) : 1292 - 1296
  • [5] BLOOD-FLOW RATE AND ACCESS RECIRCULATION IN HEMODIALYSIS
    DANIELS, ID
    BERLYNE, GM
    BARTH, RH
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1992, 15 (08) : 470 - 474
  • [6] Depner T A, 1991, ASAIO Trans, V37, pM141
  • [7] DEPNER TA, 1995, ASAIO J, V41, pM745, DOI 10.1097/00002480-199507000-00112
  • [8] DEPNER TA, 1995, ASAIO J, V41, pM749, DOI 10.1097/00002480-199507000-00113
  • [9] GRAFT GEOMETRY AND VENOUS INTIMAL-MEDIAL HYPERPLASIA IN ARTERIOVENOUS LOOP GRAFTS
    FILLINGER, MF
    REINITZ, ER
    SCHWARTZ, RA
    RESETARITS, DE
    PASKANIK, AM
    BRUCH, D
    BREDENBERG, CE
    [J]. JOURNAL OF VASCULAR SURGERY, 1990, 11 (04) : 556 - 566
  • [10] DUPLEX AND COLOR DOPPLER SONOGRAPHY OF HEMODIALYSIS ARTERIOVENOUS-FISTULAS AND GRAFTS
    FINLAY, DE
    LONGLEY, DG
    FOSHAGER, MC
    LETOURNEAU, JG
    [J]. RADIOGRAPHICS, 1993, 13 (05) : 983 - 999