In vitro comparison of frictional torque and torsional resistance of aged conventional gamma-in-nitrogen sterilized polyethylene versus aged highly crosslinked polyethylene articulating against head sizes larger than 32 mm

被引:29
作者
Burroughs, Brian R.
Muratoglu, Orhun K.
Bragdon, Charles R.
Wannomae, Keith K.
Christensen, Steven
Lozynsky, Andrew J.
Harris, William H.
机构
[1] Massachusetts Gen Hosp, Harris Orthopaed Biomech & Biomat Lab, Boston, MA 02114 USA
[2] Massachusetts Gen Hosp, Adult Reconstruct Unit, Dept Orthopaed Surg, Boston, MA 02114 USA
[3] Harvard Univ, Sch Med, Boston, MA 02114 USA
关键词
D O I
10.1080/17453670610012881
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background The advent of highly crosslinked polyethylene has allowed the re-evaluation of the use of femoral heads larger than 32 mm for metal-on-polyethylene total hip arthroplasties. However, the effect of larger heads on the frictional torque of highly crosslinked polyethylene is unknown. Methods We performed an in vitro examination of the effect of larger chrome cobalt femoral heads (40 mm diameter) on the frictional torque and torsional resistance of hip articulations on aged liners of polyethylene that were sterilized by gamma rays while in nitrogen, and aged highly crosslinked polyethylene. The frictional torque at the femoral head articulation was usually higher for the highly crosslinked polyethylene than for the conventional polyethylene. The aged conventional liners oxidized considerably, which led to gross failure of the polyethylene at the anti-rotation portion of the rim. The aged crosslinked polyethylene showed no such failures despite the higher frictional torque. Interpretation Our findings suggest that in terms of torsional resistance to fatigue when studied as a device, rather than as an isolated material, under these conditions, aged highly crosslinked polyethylene is preferable to aged conventional polyethylene.
引用
收藏
页码:710 / 718
页数:9
相关论文
共 28 条
[1]
Andersson G B, 1972, J Bone Joint Surg Br, V54, P590
[2]
[Anonymous], CLIN ORTHOPAEDICS
[3]
Compression and tension fatigue resistance of medical grade ultra high molecular weight polyethylene: the effect of morphology, sterilization, aging and temperature [J].
Baker, DA ;
Hastings, RS ;
Pruitt, L .
POLYMER, 2000, 41 (02) :795-808
[4]
Baker DA, 1999, J BIOMED MATER RES, V46, P573, DOI 10.1002/(SICI)1097-4636(19990915)46:4<573::AID-JBM16>3.0.CO
[5]
2-A
[6]
The effects of degree of crosslinking on the fatigue crack initiation and propagation resistance of orthopedic-grade polyethylene [J].
Baker, DA ;
Bellare, A ;
Pruitt, L .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2003, 66A (01) :146-154
[7]
BHATTACHARYYA S, 2004, T 50 ANN M ORTH RES
[8]
Bono J V, 1994, J Arthroplasty, V9, P119, DOI 10.1016/0883-5403(94)90059-0
[9]
BRADFORDCOLLINS L, 2002, P 28 ANN M SOC BIOM
[10]
Range of motion and stability in total hip arthroplasty with 28-, 32-, 38-, and 44-mm femoral head sizes - An in vitro study [J].
Burroughs, BR ;
Hallstrom, B ;
Golladay, GJ ;
Hoeffel, D ;
Harris, WH .
JOURNAL OF ARTHROPLASTY, 2005, 20 (01) :11-19