Initial measurements with the SUERC accelerator mass spectrometer

被引:14
作者
Freeman, S
Xu, S
Schnabel, C
Dougans, A
Tait, A
Kitchen, R
Klody, G
Loger, R
Pollock, T
Schroeder, J
Sundquist, M
机构
[1] Scottish Univ Environm Res Ctr, E Kilbride G75 0QF, Lanark, Scotland
[2] Natl Electrostat Corp, Middleton, WI USA
关键词
accelerator mass spectrometry; AMS;
D O I
10.1016/j.nimb.2004.04.040
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 [仪器科学与技术]; 080401 [精密仪器及机械]; 081102 [检测技术与自动化装置];
摘要
Be-10, C-14, Cl-36 and I-129 test measurements have been made with a new Pelletron-based accelerator mass spectrometer operating at up to 5.2 MV. All ion detection was with a versatile gas ionization detector. Low-background radiocarbon measurements with 2% scatter of identical samples was performed with both spectrometer ion sources. Be-10/Be backgrounds of 3 x 10(-15) were achieved using a gas cell adjoining the detector for B-10 suppression. High sample-throughput Cl AMS with Cl-36/Cl backgrounds of 4 x 10(-15) was accomplished. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:195 / 198
页数:4
相关论文
empty