Rotating slit image camera TOPCON SL 45 - New developments for simultaneous image acquisition by photographic and CCD-assisted on-line documentation

被引:9
作者
Dragomirescu, V [1 ]
Hockwin, O [1 ]
机构
[1] UNIV BONN,D-5300 BONN,GERMANY
关键词
slit image camera; photographic imaging; on-line imaging; combined photo and CCD imaging equipment; TOPCON SL 45 (new generation); Scheimpflug color images; lens endogenous fluorescence measurement;
D O I
10.1159/000267963
中图分类号
R77 [眼科学];
学科分类号
100212 [眼科学];
摘要
In a first approach the photographic camera part of the SL 45 B has been replaced by a CCD camera(2/3 '' black-and-white camera/Teli CS 3440 CCD B/W Tokyo Electronic Industry Co., Ltd.) unit thus enabling direct image acquisition and computer assisted analyses, Images have a full length of 8.8 mm, The images can be taken with only the observation light, no flash is necessary. Thus, the images can also be recorded on a VCR performing a complete rotation of the slit and all lens sections can be monitored obtaining almost three-dimensional information about the lens transparency/cataract formation. The further improvement is the on-line imaging for all SL-45 cameras in having CCD acquisition together with the usual 35-mm film photography, This is very important for the documentation of the lens coloration and for measuring lens fluorescence, For this TOPCON SL 45 (new generation) we are using the '2nd Scheimpflug condition' located within the eye piece of a SL 45 which is taken with a CS b/w No, 8310 C (Tokyo Electronic Industry Co., Ltd.). It is possible to integrate the CCD system in all TOPCON SL 45 Cameras (Ist and 2nd production charge as well as SL 45 B).
引用
收藏
页码:102 / 108
页数:7
相关论文
共 13 条
[1]
DRAGOMIRESCU V, 1978, INTDSCPL TOPICS GERO, V13, P118
[2]
EGNER P, 1995, INT FORUM ADV TECHNI, P40
[3]
HARRIS M L, 1991, Eye (London), V5, P487
[4]
MEASURING LENS TRANSPARENCY BY SCHEIMPFLUG PHOTOGRAPHY OF THE ANTERIOR EYE SEGMENT - PRINCIPLE, INSTRUMENTATION, AND APPLICATION TO CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY [J].
HOCKWIN, O ;
DRAGOMIRESCU, V ;
LASER, H ;
WEGENER, A ;
ECKERSKORN, U .
JOURNAL OF TOXICOLOGY-CUTANEOUS AND OCULAR TOXICOLOGY, 1987, 6 (04) :251-271
[5]
HOCKWIN O, 1990, NONINVASIVE DIAGNOST, P281
[6]
HOCKWIN O, 1995, AUGENARZTLICHE UNTER, P86
[7]
ONLINE SCHEIMPFLUG IMAGING AND ITS POTENTIAL FOR IN-VIVO EXAMINATION OF CORNEA AND LENS [J].
HUEBSCHER, HJ ;
SCHMIDT, H .
OPHTHALMIC RESEARCH, 1994, 26 :33-38
[8]
ULTRAVIOLET-VISIBLE SLIT LAMP DENSITOGRAPHY OF THE HUMAN-EYE [J].
LERMAN, S ;
HOCKWIN, O .
EXPERIMENTAL EYE RESEARCH, 1981, 33 (06) :587-596
[9]
QUANTIFICATION OF SENILE CATARACT IN THE SLIT-LAMP IMAGE OF THE ZEISS SLC MEASURING SYSTEM [J].
NIESEL, P ;
MULLER, D .
KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1988, 192 (02) :173-175
[10]
THE MULTIPURPOSE CAMERA - A NEW ANTERIOR EYE SEGMENT ANALYSIS SYSTEM [J].
SASAKI, K ;
SAKAMOTO, Y ;
SHIBATA, T ;
EMORI, Y .
OPHTHALMIC RESEARCH, 1990, 22 :3-8