Incidence of Ventricular Arrhythmias in Patients on Long-term Support With a Continuous-flow Assist Device (HeartMate II)

被引:131
作者
Andersen, Mads [1 ]
Videbaek, Regitze [1 ]
Boesgaard, Soren [1 ]
Sander, Kare [2 ]
Hartsen, Peter B. [3 ]
Gustafsson, Finn [1 ]
机构
[1] Rigshosp, Ctr Heart, Dept Cardiol, DK-2100 Copenhagen, Denmark
[2] Rigshosp, Ctr Heart, Dept Cardiothorac Surg, DK-2100 Copenhagen, Denmark
[3] Rigshosp, Ctr Heart, Dept Cardiothorac Anesthesia, DK-2100 Copenhagen, Denmark
关键词
FAILURE;
D O I
10.1016/j.healun.2009.03.011
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The incidence of ventricular tachycardia (VT) or ventricular fibrillation (VF) in patients supported with a continuous-flow left ventricular assist device (LVAD) has not been investigated in detail. in 23 consecutive recipients of a HeartMate II, we analyzed the incidence of VT/VF during a total of 266 months of follow-up. Sustained VT or VF occurred in 52% of the patients, with the majority of arrhythmias occurring in the first 4 weeks after LVAD implantation. VT/VF requiring implantable cardioverter-defibrillator (ICD) shock or external defibrillation occurred in 8 patients and significant hemodynamic instability ensued ill 3 patients. There were no clear predictors of VT/VF, and it is argued that prophylactic ICD implantation should be considered in patients supported with a continuous-flow LVAD. J Heart Lung Transplant 2009;28:733-5. Copyright (C) 2009 by the International Society for Heart and Lung Transplantation.
引用
收藏
页码:733 / 735
页数:3
相关论文
共 9 条
[1]  
Arai H, 1991, ASAIO Trans, V37, pM427
[2]   Ventricular arrhythmias during left ventricular assist device support [J].
Bedi, Maninder ;
Kormos, Robert ;
Winowich, Steve ;
McNamara, Dennis M. ;
Mathier, Michael A. ;
Murali, Srinivas .
AMERICAN JOURNAL OF CARDIOLOGY, 2007, 99 (08) :1151-1153
[3]   Left ventricular assist device and drug therapy for the reversal of heart failure [J].
Birks, Emma J. ;
Tansley, Patrick D. ;
Hardy, James ;
George, Robert S. ;
Bowles, Christopher T. ;
Burke, Margaret ;
Banner, Nicholas R. ;
Khaghani, Asghar ;
Yacoub, Magdi H. .
NEW ENGLAND JOURNAL OF MEDICINE, 2006, 355 (18) :1873-1884
[4]   Prolonged repolarization after ventricular assist device support is associated with arrhythmias in humans with congestive heart failure [J].
Harding, JD ;
Piacentino, V ;
Rothman, S ;
Chambers, S ;
Jessup, M ;
Margulies, KB .
JOURNAL OF CARDIAC FAILURE, 2005, 11 (03) :227-232
[5]   Use of a continuous-flow device in patients awaiting heart transplantation [J].
Miller, Leslie W. ;
Pagani, Francis D. ;
Russell, Stuart D. ;
John, Ranjit ;
Boyle, Andrew J. ;
Aaronson, Keith D. ;
Conte, John V. ;
Naka, Yoshifumi ;
Mancini, Donna ;
Delgado, Reynolds M. ;
MacGillivray, Thomas E. ;
Farrar, David J. ;
Frazier, O. H. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (09) :885-896
[6]   Ventricular arrhythmias after left ventricular assist device implantation [J].
Refaat, Marwan ;
Chemaly, Elie ;
Lebeche, Djamel ;
Gwathmey, Judith K. ;
Hajjar, Roger J. .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2008, 31 (10) :1246-1252
[7]  
Rohit Mehta, 2007, J Cardiovasc Electrophysiol, V18, pE28, DOI 10.1111/j.1540-8167.2007.00949.x
[8]   HeartMate II left ventricular assist device;: early European experience [J].
Strueber, Martin ;
Sander, Kare ;
Lahpor, Jaap ;
Ahn, Henrik ;
Litzler, Pierre-Yves ;
Drakos, Stavros G. ;
Musumeci, Francesco ;
Schtensak, Christian ;
Friedrich, Ivar ;
Gustafsson, Ronny ;
Oertel, Frank ;
Leprince, Pascal .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2008, 34 (02) :289-294
[9]   Suction events during left ventricular support and ventricular arrhythmias [J].
Vollkron, Michael ;
Voitl, Peter ;
Ta, Julia ;
Wieselthaler, Georg ;
Schima, Heinrich .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2007, 26 (08) :819-825