Optimization of an HPLC-HG-AFS method for screening Sb(V), Sb(III), and Me3SbBr2 in water samples

被引:29
作者
Sayago, A [1 ]
Beltrán, R [1 ]
Recamales, MAF [1 ]
Gómez-Ariza, JL [1 ]
机构
[1] Escuela Politecn Super, Dept Quim & Ciencia Mat, Palos De La Frontera 21819, Huelva, Spain
关键词
D O I
10.1039/b202065n
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A procedure for the separation and determination of inorganic and organic species of antimony based on the use of coupled high- performance liquid chromatography ( HPLC) hydride generation ( HG) atomic fluorescence spectrometry ( AFS) is described. Me3SbBr2 and inorganic Sb( V) and Sb( III) were separated on an anion-exchange column [Hamilton PRPX- 100 ( 100 x 4.1 mm)] using the concentration gradient between potassium hydroxide, 20 mM pH 11, and ammonium tartrate, 200 mM pH 5. The flow rate through the column was 1.5 mL min(-1). The retention times were 0.98, 1.71 and 5.03 min for Me3SbBr2, Sb( V) and Sb( III), respectively. The hydrides were generated on- line with 3% m/v NaBH4 and 2 mol l(-1) HNO3 solutions. The intensity of emission was linearly related to the concentration of the derivatized Me3SbBr2 species in the range 1.0-50.0 mug l(-1) ( r = 0.9996), Sb(v) species in the range 1.0-50.0 mug l(-1) ( r = 0.9998) and Sb(III) in the range 2.5-50.0 mug l(-1) ( r = 0.9998), with detection limits of 0.04 mug l(-1) for Me3SbBr2, 0.09 mug l(-1) for Sb( V) and 0.26 mug l(-1) for Sb( III). The precisions evaluated using the relative standard deviation ( RSD) were 6.8%, 2.4% and 3.3% for Me3SbBr2, Sb( V) and Sb( III), respectively. The method has sensitivity suitable for monitoring drinking and surface waters according to the regulations established by the EU directives for antimony (10 mug l(-1)), and it was applied to the speciation of inorganic and organic antimony in several spiked water samples. The precisions for well water, Merck and mineral water were limited in relation to Me3SbBr2 due to its conversion to Sb( V).
引用
收藏
页码:1400 / 1404
页数:5
相关论文
共 21 条
[1]  
Andrewes P, 1998, APPL ORGANOMET CHEM, V12, P827, DOI 10.1002/(SICI)1099-0739(199812)12:12<827::AID-AOC797>3.3.CO
[2]  
2-F
[3]  
[Anonymous], 1989, TRACE ELEMENTS HUMAN
[4]   Development of a multi-element hydride generation inductively coupled plasma mass spectrometry procedure for the simultaneous determination of arsenic, antimony and selenium in waters [J].
Bowman, J ;
Fairman, B ;
Catterick, T .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (03) :313-316
[5]   DIRECT VOLATILIZATION SPECTRAL EMISSION TYPE DETECTION SYSTEM FOR NANOGRAM AMOUNTS OF ARSENIC AND ANTIMONY [J].
BRAMAN, RS ;
FOREBACK, CC ;
JUSTEN, LL .
ANALYTICAL CHEMISTRY, 1972, 44 (13) :2195-&
[6]   APPLICABILITY OF ANODIC-STRIPPING VOLTAMMETRY AND GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY TO THE DETERMINATION OF ANTIMONY IN BIOLOGICAL MATRICES - A COMPARATIVE-STUDY [J].
COSTANTINI, S ;
GIORDANO, R ;
RIZZICA, M ;
BENEDETTI, F .
ANALYST, 1985, 110 (11) :1355-1359
[7]  
DEDINA J, 1995, HYDRIDE GENERATION A
[8]   METHYLATED ANTIMONY(V) COMPOUNDS - SYNTHESIS, HYDRIDE GENERATION PROPERTIES AND IMPLICATIONS FOR AQUATIC SPECIATION [J].
DODD, M ;
GRUNDY, SL ;
REIMER, KJ ;
CULLEN, WR .
APPLIED ORGANOMETALLIC CHEMISTRY, 1992, 6 (02) :207-211
[9]  
DULIVO A, 1995, J ANAL ATOM SPECTROM, V10, P696
[10]   Determination of arsenic, antimony, selenium, tellurium and bismuth in nickel metal by hydride generation atomic fluorescence spectrometry [J].
Feng, XJ ;
Fu, B .
ANALYTICA CHIMICA ACTA, 1998, 371 (01) :109-113