A micro-XRF spectrometer based on a rotating anode generator and capillary optics

被引:47
作者
Janssens, K
Vekemans, B
Vincze, L
Adams, F
Rindby, A
机构
[1] Department of Chemistry, University of Antwerp (UIA)
[2] Department of Physics, Chalmers University of Technology
关键词
capillary optics; corrosion; glass; microanalysis; micro-X-ray fluorescence spectrometer; porcelain; trace element distributions; X-ray fluorescence;
D O I
10.1016/S0584-8547(96)01551-0
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
The analytical characteristics of a laboratory-scale micro-X-ray fluorescence spectrometer, based on a rotating anode X-ray generator and capillary optics, are described. Usually, a microbeam 15 mu m in diameter, derived from a copper or molybdenum anode operated at 45 kV, is used for sample irradiation. Elemental yields around the 1 count s(-1) (mu g cm(-2))(-1) level are obtained, corresponding to absolute detection limits for thin samples in the 0.05-1 pg range and to relative MDL levels of 3-10 ppm for thick organic samples. The use of the instrument for studying (trace) element migration in Roman glass and for the nondestructive analysis of decorations on Japanese I8th century porcelain vases is described.
引用
收藏
页码:1661 / 1678
页数:18
相关论文
共 36 条
[1]  
AERTS A, 1994, P 28 MAS M NEW ORL L, P45
[2]   IMPROVED CAPILLARY OPTICS APPLIED TO MICROBEAM X-RAY-FLUORESCENCE - RESOLUTION AND SENSITIVITY [J].
ATTAELMANAN, A ;
VOGLIS, P ;
RINDBY, A ;
LARSSON, S ;
ENGSTROM, P .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (01) :24-27
[3]  
BILDERBACK D, 1996, COMMUNICATION
[4]  
BOEHME DR, 1987, ADV XRAY ANAL, V30, P39
[5]   SYNCHROTRON RADIATION-INDUCED X-RAY-FLUORESCENCE AT LURE [J].
BRISSAUD, I ;
WANG, JX ;
CHEVALLIER, P .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1989, 131 (02) :399-413
[6]  
CARPENTER D, 1994, P 28 MAS M NEW ORL L, P17
[7]  
CARPENTER DA, 1991, ADV X RAY ANAL, V34, P217
[8]  
CARPENTER DA, 1995, J TRACE MICROPROBE T, V13, P141
[9]  
CARPENTER DA, 1989, ADV XRAY ANAL, V32, P115
[10]   MODELING THE TRANSMISSION OF X-RAYS THROUGH CLASS CAPILLARY WAVE-GUIDES - IMPLICATIONS FOR THE DESIGN OF A LABORATORY X-RAY MICROPROBE [J].
CHARNLEY, NR ;
POTTS, PJ ;
LONG, JVP .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (11) :1185-1193