Robustness testing of a reversed-phase high-performance liquid chromatographic assay: comparison of fractional and asymmetrical factorial designs

被引:47
作者
Hund, E
Vander Heyden, Y
Haustein, M
Massart, DL
Smeyers-Verbeke, J
机构
[1] Free Univ Brussels, Fannaceutisch Inst, B-1090 Brussels, Belgium
[2] Bayer AG, Zentral Analyt Dormagen, D-41538 Dormagen, Germany
关键词
robustness test; fractional factorial design; half-normal plots; critical effects;
D O I
10.1016/S0021-9673(00)00081-9
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Robustness tests were performed on a reversed-phase HPLC assay for triadimenol. Different experimental designs were compared. Two-level fractional factorial designs with different resolutions were used to study the influence of procedure-related factors. The factors chromatographic column manufacturer at four levels and instrument at three levels were stepwise included in the study using asymmetrical factorial designs. The significance of the factor effects was determined statistically, using two types of error estimates in the calculation of critical effects, and graphically, by means of half-normal plots. The asymmetrical designs turned out to be an efficient and economic method to examine the influence of factors at different numbers of levels in the robustness testing of analytical methods. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:167 / 185
页数:19
相关论文
共 25 条
[1]   ORTHOGONAL MAIN-EFFECT PLANS FOR ASYMMETRICAL FACTORIAL EXPERIMENTS [J].
ADDELMAN, S .
TECHNOMETRICS, 1962, 4 (01) :21-+
[2]   Usefulness of a ruggedness test in the validation of flow injection analysis systems [J].
Alvares-Ribeiro, LMBC ;
Machado, AASC .
ANALYTICA CHIMICA ACTA, 1997, 355 (2-3) :195-201
[3]  
[Anonymous], DESIGN OPTIMISATION
[4]  
[Anonymous], STP PHARM PRATIQUES
[5]  
[Anonymous], 1996, INT C HARM TECHN REQ
[6]  
Daniel C., 1959, TECHNOMETRICS, V1, P311, DOI [10.1080/00401706.1959.10489866, DOI 10.1080/00401706.1959.10489866]
[7]  
DONG F, 1993, STAT SINICA, V3, P209
[8]  
DYSON N, 1990, CHROMATOGRAPHIC INTE, P35
[10]   Outliers and experimental designs [J].
Goupy, J .