Accuracy of results obtained by performing a second ligase chain reaction assay and PCR analysis on urine samples with positive or near-cutoff results in the LCx test for Chlamydia trachomatis

被引:19
作者
Castriciano, S
Luinstra, K
Jang, D
Patel, J
Mahony, J
Kapala, J
Chernesky, M
机构
[1] McMaster Univ, St Josephs Healthcare, Father Sean OSullivan Res Ctr, Hamilton, ON L8N 4A6, Canada
[2] McMaster Univ, Hamilton Reg Lab, Program Med, Hamilton, ON L8N 4A6, Canada
[3] McMaster Univ, Dept Pathol, Hamilton, ON L8N 4A6, Canada
[4] McMaster Univ, Dept Mol Med, Hamilton, ON L8N 4A6, Canada
[5] Dept Microbiol, Gamma Dynacare Med Labs, Brampton, ON L6T 5M3, Canada
关键词
D O I
10.1128/JCM.40.7.2632-2634.2002
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Nucleic acid amplification assays such as the ligase chain reaction and PCR have encountered reproducibiIity problems. The initial extract and a newly extracted aliquot of urine specimens (n = 120) which had signaI-to-cutoff (S/CO) ratios above 0.80 by the LCx Chlamydia assay were retested. Nucleic acid was extracted from an additional urine sample for testing by the AMPLICOR PCR Chlamydia assay. Fifteen percent (18 of 120) of the urine specimens were negative by all repeat tests (initial mean S/CO ratio by the LCx Chlaniydia assay, 0.93; S/CO ratio range, 0.80 to 3.30). Repeat testing of the 102 specimens with possible positive results by the LCx Chlamydia assay by use of the initially extracted aliquot confirmed the results for 95 (93.1%) of the specimens; repeat testing of a newly extracted aliquot confirmed the results for 87 (85.3%) of the specimens. Twenty specimens had discordant results by the two repeat LCx Chlamydia assays. A total of 78 of 102 (76.5%) of the specimens were positive by the AMPLICOR PCR, and the AMPLICOR PCR confirmed the results for 82.1% (78 of 95) and 89.6% (78 of 87) of the specimens positive by the two repeat LCx Chlamydia assays, respectively. Some of the discrepancies observed by multiple repeat tests may have been due to specimen mislabeling or contamination during performance of the procedure rather than to the LCx Chlamydia assay. Both assays suffered from a lack of reproducibility on repeat testing with a small proportion of specimens, probably due to the presence of low levels of DNA, the presence of variable amounts of amplification inhibitors, and the loss of DNA during extraction.
引用
收藏
页码:2632 / 2634
页数:3
相关论文
共 10 条
  • [1] DIAGNOSIS OF CHLAMYDIA-TRACHOMATIS URETHRITIS IN MEN BY POLYMERASE CHAIN-REACTION ASSAY OF 1ST-CATCH URINE
    BAUWENS, JE
    CLARK, AM
    LOEFFELHOLZ, MJ
    HERMAN, SA
    STAMM, WE
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1993, 31 (11) : 3013 - 3016
  • [2] DIAGNOSIS OF CHLAMYDIA-TRACHOMATIS INFECTIONS IN MEN AND WOMEN BY TESTING FIRST-VOID URINE BY LIGASE CHAIN-REACTION
    CHERNESKY, MA
    JANG, D
    LEE, H
    BURCZAK, JD
    HU, H
    SELLORS, J
    TOMAZICALLEN, SJ
    MAHONY, JB
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1994, 32 (11) : 2682 - 2685
  • [3] Coutlée F, 2000, J CLIN MICROBIOL, V38, P2512
  • [4] Gronowski AM, 2000, J CLIN MICROBIOL, V38, P2416
  • [5] DIRECT DETECTION OF CHLAMYDIA-TRACHOMATIS IN URINE SPECIMENS FROM SYMPTOMATIC AND ASYMPTOMATIC MEN BY USING A RAPID POLYMERASE CHAIN-REACTION ASSAY
    JASCHEK, G
    GAYDOS, CA
    WELSH, LE
    QUINN, TC
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1993, 31 (05) : 1209 - 1212
  • [6] Urine specimens from pregnant and nonpregnant women inhibitory to amplification of Chlamydia trachomatis nucleic acid by PCR, ligase chain reaction, and transcription-mediated amplification:: Identification of urinary substances associated with inhibition and removal of inhibitory activity
    Mahony, J
    Chong, S
    Jang, D
    Luinstra, K
    Faught, M
    Dalby, D
    Sellors, J
    Chernesky, M
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1998, 36 (11) : 3122 - 3126
  • [7] Reproducibility problems with the AMPLICOR PCR Chlamydia trachomatis test
    Peterson, EM
    Darrow, V
    Blanding, J
    Aarnaes, S
    delaMaza, LM
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1997, 35 (04) : 957 - 959
  • [8] LIGASE CHAIN-REACTION TO DETECT CHLAMYDIA-TRACHOMATIS INFECTION OF THE CERVIX
    SCHACHTER, J
    STAMM, WE
    QUINN, TC
    ANDREWS, WW
    BURCZAK, JD
    LEE, HH
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1994, 32 (10) : 2540 - 2543
  • [9] Replicate PCR testing and probit analysis for detection and quantitation of Chlamydia pneumoniae in clinical specimens
    Smieja, M
    Mahony, JB
    Goldsmith, CH
    Chong, S
    Petrich, A
    Chernesky, M
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2001, 39 (05) : 1796 - 1801
  • [10] Performance of transcription-mediated amplification and ligase chain reaction assays for detection of chlamydial infection in urogenital samples obtained by invasive and noninvasive methods
    Stary, A
    Schuh, E
    Kerschbaumer, M
    Götz, B
    Lee, H
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1998, 36 (09) : 2666 - 2670