Simultaneous determination of triamcinolone acetonide and hydrocortisone in human plasma by high-performance liquid chromatography

被引:33
作者
Doppenschmitt, SA
Scheidel, B
Harrison, F
Surmann, JP
机构
[1] HARRISON CLIN RES GMBH,D-80636 MUNICH,GERMANY
[2] HUMBOLDT UNIV BERLIN,MAT NAT FAK 1,INST PHARM,D-13363 BERLIN,GERMANY
来源
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS | 1996年 / 682卷 / 01期
关键词
triamcinolone acetonide; hydrocortisone; corticoids;
D O I
10.1016/0378-4347(96)00060-6
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A validated HPLC method for the simultaneous determination of triamcinolone acetonide and hydrocortisone has been established to monitor the plasma levels of the compounds in healthy volunteers following intramuscular (i.m.) administration of triamcinolone acetonide. Plasma (1.0 ml) was extracted with dichloromethane after addition of the internal standard, fluocortolone. The compounds were separated using a LiChrospher RP 18 column and detected by UV absorbance. Specificity, linearity, as well as the repeatability, intermediate precision and accuracy of the method were established. The limit of quantification was 0.6 ng/ml for triamcinolone acetonide (C.V. = 8.7%, R.E. = 2.6%, n = 6) and 2.0 ng/ml for hydrocortisone (C.V. = 8.3%, R.E. = 2.8%, n = 6). Data on the stability of triamcinolone acetonide in human plasma are presented. Recovery of the compounds and the internal standard have been studied. The results of quality control samples (n = 126) determined during routine analysis of volunteer samples are described. Plasma levels of triamcinolone acetonide after i.m. administration of 40 mg of triamcinolone acetonide are presented.
引用
收藏
页码:79 / 88
页数:10
相关论文
共 23 条
[1]   EFFECT OF EXPERIMENTAL-DESIGN ON ASSAY CALIBRATION [J].
AARONS, L .
ANALYST, 1981, 106 (1269) :1249-1254
[2]  
BLAUERTCOUSOUNIS SP, 1989, J ALLERGY CLIN IMMUN, V83, P221
[3]   PRECISION-DEPENDENT CALIBRATION IN INSTRUMENTAL ANALYSIS [J].
BUBERT, H ;
KLOCKENKAMPER, R .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1983, 316 (02) :186-193
[4]   METHOD VALIDATION IN THE BIOANALYTICAL LABORATORY [J].
BUICK, AR ;
DOIG, MV ;
JEAL, SC ;
LAND, GS ;
MCDOWALL, RD .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1990, 8 (8-12) :629-637
[5]  
Cartwright A, 1991, DRUG INF J, V25, P471, DOI [10.1177/009286159102500402, DOI 10.1177/009286159102500402]
[6]   EVALUATION OF CRITERIA FOR THE ACCEPTANCE OF BIOANALYTICAL DATA [J].
CAUSEY, AG ;
HILL, HM ;
PHILLIPS, LJ .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1990, 8 (8-12) :625-628
[7]   PHARMACOKINETICS AND PHARMACODYNAMICS OF GLUCOCORTICOID SUSPENSIONS AFTER INTRAARTICULAR ADMINISTRATION [J].
DERENDORF, H ;
MOLLMANN, H ;
GRUNER, A ;
HAACK, D ;
GYSELBY, G .
CLINICAL PHARMACOLOGY & THERAPEUTICS, 1986, 39 (03) :313-317
[8]  
DERENDORF H, 1986, J PHARM BIOMED ANAL, V2, P197
[9]   NON-CONSTANT VARIANCE REGRESSION TECHNIQUES FOR CALIBRATION-CURVE-BASED ANALYSIS [J].
GARDEN, JS ;
MITCHELL, DG ;
MILLS, WN .
ANALYTICAL CHEMISTRY, 1980, 52 (14) :2310-2315
[10]  
GARDNER RS, 1990, J PHARM BIOMED ANAL, V8, P8