Design of superconducting MRI surface coil by using method of moment

被引:20
作者
Fang, J [1 ]
Chow, MS
Chan, KC
Wong, KK
Shen, GX
Gao, E
Yang, ES
Ma, QY
机构
[1] Univ Hong Kong, Jockey Club MRI Engn Ctr, Hong Kong, Hong Kong, Peoples R China
[2] Univ Hong Kong, Dept Elect & Elect Engn, Hong Kong, Hong Kong, Peoples R China
[3] Columbia Univ, Dept Elect Engn, New York, NY 10027 USA
关键词
MoM; MRI; mutual coupling; RF coil;
D O I
10.1109/TASC.2002.1020344
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A method of moment with. an enhanced model to design high-temperature superconductor (HTS) RF surface coils for magnetic resonant image (MRI) is presented. The resonant frequency and quality factor (Q) of HTS RF spiral coils are simulated using this method. The agreements of resonant frequencies and Qs between the simulation and measurement are excellent with differences,less than 1% and 3%, respectively. The 0.2-mu m-thick YBaCuO.(YBCO) thin films are deposited onto single side of 0.508-mm-thick LaAlO3 (LAO) and sapphire substrate and patterned into a spiral shape. To accurately analyze the resonant frequency and Q of a coil, an enhanced two-fluid model is employed. HTS RF coils with diameter of 65 mm for 0.2T and 1.5T MRI systems are designed and fabricated with the measured Q of 19 K and 23 K, respectively. In addition, the shift of resonant frequency due to. the mutual coupling between two HTS spiral coils is predicted by this method, which is important for design of HTS coil arrays in an MRI system.
引用
收藏
页码:1823 / 1827
页数:5
相关论文
共 13 条
[1]   Circuit properties of coils [J].
Anicin, BA ;
Davidovic, DM ;
Karanovic, P ;
Miljevic, VM ;
Radojevic, V .
IEE PROCEEDINGS-SCIENCE MEASUREMENT AND TECHNOLOGY, 1997, 144 (05) :234-239
[2]   NUCLEAR MAGNETIC-RESONANCE MICROSCOPY WITH 4-MU-M RESOLUTION - THEORETICAL-STUDY AND EXPERIMENTAL RESULTS [J].
CHO, ZH ;
AHN, CB ;
JUH, SC ;
LEE, HK ;
JACOBS, RE ;
LEE, S ;
YI, JH ;
JO, JM .
MEDICAL PHYSICS, 1988, 15 (06) :815-824
[3]  
Gupta KC, 1981, COMPUTER AIDED DESIG
[4]   Accurate distributed inductance of spiral resonators [J].
Hejazi, ZM ;
Excell, PS ;
Jiang, Z .
IEEE MICROWAVE AND GUIDED WAVE LETTERS, 1998, 8 (04) :164-166
[5]   SENSITIVITY OF THE ZEUGMATOGRAPHIC EXPERIMENT INVOLVING HUMAN SAMPLES [J].
HOULT, DI ;
LAUTERBUR, PC .
JOURNAL OF MAGNETIC RESONANCE, 1979, 34 (02) :425-433
[6]  
*HP EES, HP MOM US GUID
[7]   Calculation of distributed capacitances of spiral resonators [J].
Jiang, ZF ;
Excell, PS ;
Hejazi, ZM .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1997, 45 (01) :139-142
[8]   MRI simulation-based evaluation of image-processing and classification methods [J].
Kwan, RKS ;
Evans, AC ;
Pike, GB .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1999, 18 (11) :1085-1097
[9]   Electromagnetics in high-T-c superconductors [J].
Ma, JG ;
Wolff, I .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1996, 44 (04) :537-542
[10]   RF applications of high temperature superconductors in MHz range [J].
Ma, QY .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1999, 9 (02) :3565-3568