High speed photographic study of wet droplets and solid particles in the inductively coupled plasma

被引:56
作者
Houk, RS
Winge, RK
Chen, XS
机构
[1] Ames Laboratory, US Department of Energy, Iowa State University, Ames
[2] Metropol. Water Dist. S. California, La Verne, CA 91750-3399
关键词
inductively coupled plasma; inductively coupled plasma mass spectrometry; nebulization; sample introduction; slurry; noise; particles; droplets;
D O I
10.1039/a607579g
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Motion pictures of the ICP were taken at 4000 frames s(-1), An intact wet droplet containing yttrium causes the formation of a pale red cloud juxtaposed on the usual ambient emission structure of the plasma, Most of these droplet clouds are shaped like an oval or a comet, A few droplets produce small, bright spheres followed by faint, wispy streaks that point downstream, Such a spherical cloud is caused by some rapid event such as explosion of a droplet in the final stages of solvent evaporation, The faint sneaks are some residue, perhaps small solid particles, In a particular frame, a number of these faint streaks protrude from the tip of the initial radiation zone (IRZ) into tbe normal analytical zone (NAZ). When wet droplets are introduced, that portion of the analyte that navels through the center of the plasma passes through three distinct regions (i.e., the IRZ, the streaks, then the NAZ), rather than directly from the IRZ to the NAZ, Groups of two or three droplets tend to appear together in the same time interval (approximate to 0.12 ms) in the plasma, These droplet clouds are not seen, and the particle streaks are much less evident, when the solvent is removed before the aerosol is injected into the plasma, Aqueous slurries of Y2O3 in various particle sizes (0.1 or 3 mu m mean diameter) produce white streaks along the center line of the plasma, which are attributed to individual solid particles, These observations also support the general precept that calibration of the response for slurries using aqueous solution standards is best accomplished by keeping the particle loading such that each wet droplet contains no more than one solid slurry particle.
引用
收藏
页码:1139 / 1148
页数:10
相关论文
共 45 条
[1]   MASS-SPECTRAL INVESTIGATION OF MECHANISMS OF LEAD VAPORIZATION FROM A GRAPHITE SURFACE USED IN ELECTROTHERMAL ATOMIZERS [J].
BASS, DA ;
HOLCOMBE, JA .
ANALYTICAL CHEMISTRY, 1987, 59 (07) :974-980
[2]   INVESTIGATION INTO VAPORIZATION OF INDIVIDUAL SOLUTE PARTICLES IN FLAME SPECTROMETRY [J].
BASTIAAN.GJ ;
HIEFTJE, GM .
ANALYTICAL CHEMISTRY, 1974, 46 (07) :901-910
[3]   MORPHOLOGICAL INVESTIGATION OF SOLUTE PARTICLES TRAPPED FROM AN AIR-ACETYLENE FLAME [J].
CHILDERS, AG ;
HIEFTJE, GM .
APPLIED SPECTROSCOPY, 1986, 40 (07) :939-944
[4]   A SIMPLE, NON-INVASIVE METHOD FOR THE MEASUREMENT OF GAS-FLOW VELOCITIES IN THE INDUCTIVELY COUPLED PLASMA [J].
CICERONE, MT ;
FARNSWORTH, PB .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1989, 44 (09) :897-907
[5]   DIAGNOSTIC STUDIES ON DESOLVATED AEROSOLS FROM ULTRASONIC NEBULIZERS [J].
CLIFFORD, RH ;
SOHAL, P ;
LIU, HY ;
MONTASER, A .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1992, 47 (09) :1107-1122
[6]  
DIEHL H, 1952, QUANTITATIVE ANAL, P32
[7]   Time resolved inductively coupled plasma mass spectrometry measurements with individual, monodisperse drop sample introduction [J].
Dziewatkoski, MP ;
Daniels, LB ;
Olesik, JW .
ANALYTICAL CHEMISTRY, 1996, 68 (07) :1101-1109
[8]   DIRECT ATOMIC SPECTROMETRIC ANALYSIS BY SLURRY ATOMIZATION .9. FUNDAMENTAL-STUDIES OF REFRACTORY SAMPLES [J].
EBDON, L ;
FOULKES, ME ;
HILL, S .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1990, 5 (01) :67-73
[9]   ULTRASONIC NEBULIZATION OF LIQUID SAMPLES FOR ANALYTICAL INDUCTIVELY COUPLED PLASMA-ATOMIC SPECTROSCOPY - AN UPDATE [J].
FASSEL, VA ;
BEAR, BR .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1986, 41 (10) :1089-1113
[10]   MONODISPERSE DRIED MICROPARTICULATE INJECTOR FOR ANALYTICAL INSTRUMENTATION [J].
FRENCH, JB ;
ETKIN, B ;
JONG, R .
ANALYTICAL CHEMISTRY, 1994, 66 (05) :685-691