Higher throughput bioanalysis by automation of a protein precipitation assay using a 96-well format with detection by LC-MS/MS

被引:97
作者
Watt, AP [1 ]
Morrison, D [1 ]
Locker, KL [1 ]
Evans, DC [1 ]
机构
[1] Merck Sharp & Dohme Res Labs, Neurosci Res Ctr, Dept Med Chem, Drug Metab Sect, Harlow CM20 2QR, Essex, England
关键词
D O I
10.1021/ac9906633
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Generic methodology for the automated preparation and analysis of drug levels in plasma samples within a drug discovery environment was achieved through the redesign of a protein precipitation assay to a microtiter (96-well) plate format and the application of robotic liquid handling for performance of all transfer and pipetting steps. Validation studies revealed that the application of robotics to sample preparation, in general, maintained the analytical accuracy and precision compared with preparing samples manually. The use of rapid gradient LC-MS/MS for analysis coupled with flow diversion of the solvent front allowed the introduction of protein-precipitated samples into the mass spectrometer without the necessity for source cleaning. The problem inherent in automatically pipetting plasma, caused by fibrinogen clots, was overcome by storing samples at -80 degrees C and thus precluding clot formation. The resulting methodology allowed sample preparation for a 96-well plate designed to accommodate 54 unknowns, duplicate 12-point calibration curves, and 6 sets of quality controls at three levels in approximately 2 h, This approach allowed an increase in throughput of sample preparation and analysis to >400 samples per day per LC-MS/MS instrument with minimal manual intervention. Overall, substantial time savings were realized, demonstrating that automation is an increasingly essential tool in a drug discovery bioanalytical environment.
引用
收藏
页码:979 / 984
页数:6
相关论文
共 37 条
  • [1] Allanson JP, 1996, RAPID COMMUN MASS SP, V10, P811, DOI 10.1002/(SICI)1097-0231(199605)10:7<811::AID-RCM561>3.0.CO
  • [2] 2-Q
  • [3] Optimisation and routine use of generic ultra-high flow-rate liquid chromatography with mass spectrometric detection for the direct on-line analysis of pharmaceuticals in plasma
    Ayrton, J
    Dear, GJ
    Leavens, WJ
    Mallett, DN
    Plumb, RS
    [J]. JOURNAL OF CHROMATOGRAPHY A, 1998, 828 (1-2) : 199 - 207
  • [4] Ayrton J, 1997, RAPID COMMUN MASS SP, V11, P1953, DOI 10.1002/(SICI)1097-0231(199712)11:18<1953::AID-RCM102>3.0.CO
  • [5] 2-Z
  • [6] Bayliss M K, 1999, Curr Opin Drug Discov Devel, V2, P20
  • [7] Beaudry F, 1998, RAPID COMMUN MASS SP, V12, P1216, DOI 10.1002/(SICI)1097-0231(19980915)12:17<1216::AID-RCM304>3.0.CO
  • [8] 2-O
  • [9] AUTOMATIC SOLID-PHASE EXTRACTION AND HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF QUINIDINE IN PLASMA
    BRANDSTETEROVA, E
    ROMANOVA, D
    KRALIKOVA, D
    BOZEKOVA, L
    KRISKA, M
    [J]. JOURNAL OF CHROMATOGRAPHY A, 1994, 665 (01) : 101 - 104
  • [10] Broach JR, 1996, NATURE, V384, P14