A micropower dry-electrode ECG preamplifier

被引:85
作者
Burke, MJ [1 ]
Gleeson, DT [1 ]
机构
[1] Univ Dublin Trinity Coll, Dept Elect & Elect Engn, Dublin 2, Ireland
关键词
dry electrodes; electrocardiogram (ECG) amplifiers; ECG recording; instrumentation amplifiers; low power;
D O I
10.1109/10.821734
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper describes the development of a very low-power preamplifier intended for use in pasteless-electrode recording of the human electrocardiogram, The expected input signal range is 100 mu V-10 mV from a lead-II electrode configuration. The amplifier provides a gain of 43 dB in a 3-dB bandwidth of 0.05 Hz-2 kHz with a defined high input impedance of 75 M Omega. It uses a driven common electrode to enhance rejection of common-mode interfering signals, including low-frequencg motion artifact, achieving a common-mode rejection ratio (CMRR) of better than 80 dB over its entire bandwidth, The gain and phase characteristics meet the recommendations of the American Heart Association, ensuring low distortion of the output ECG signal and making it suitable for clinical monitoring. The amplifier has a power consumption of 30 mu W operating from a 3.3-V battery and is intended for use in small, lightweight, portable electrocardiographic equipment and heart-rate monitoring instrumentation.
引用
收藏
页码:155 / 162
页数:8
相关论文
共 24 条
[1]   RECOMMENDATIONS FOR STANDARDIZATION AND SPECIFICATIONS IN AUTOMATED ELECTROCARDIOGRAPHY - BANDWIDTH AND DIGITAL SIGNAL-PROCESSING - A REPORT FOR HEALTH-PROFESSIONALS BY AN AD HOC WRITING GROUP OF THE COMMITTEE ON ELECTROCARDIOGRAPHY AND CARDIAC ELECTROPHYSIOLOGY OF THE COUNCIL-ON-CLINICAL-CARDIOLOGY, AMERICAN-HEART-ASSOCIATION [J].
BAILEY, JJ ;
BERSON, AS ;
GARSON, A ;
HORAN, LG ;
MACFARLANE, PW ;
MORTARA, DW ;
ZYWIETZ, C .
CIRCULATION, 1990, 81 (02) :730-739
[2]   ELECTROCARDIOGRAM RECORDING WITH PASTELESS ELECTRODES [J].
BERGEY, GE ;
SQUIRES, RD ;
SIPPLE, WC .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1971, BM18 (03) :206-&
[3]   ELECTROCARDIOGRAPHIC DISTORTIONS CAUSED BY INADEQUATE HIGH-FREQUENCY RESPONSE OF DIRECT-WRITING ELECTROCARDIOGRAPHS [J].
BERSON, AS ;
PIPBERGER, HV .
AMERICAN HEART JOURNAL, 1967, 74 (02) :208-+
[4]   LOW-FREQUENCY RESPONSE OF ELECTROCARDIOGRAPHS A FREQUENT SOURCE OF RECORDING ERRORS [J].
BERSON, AS ;
PIPBERGER, HV .
AMERICAN HEART JOURNAL, 1966, 71 (06) :779-+
[5]  
Brouha L., 1943, REV CANA DIENNE BIOL, V2, P86
[6]  
BURKE MJ, 1990, THESIS TRINITY COLLE
[7]   INSULATED ELECTROCARDIOGRAM ELECTRODES [J].
DAVID, RM ;
PORTNOY, WM .
MEDICAL & BIOLOGICAL ENGINEERING, 1972, 10 (06) :742-751
[8]  
Gatzke R., 1974, BIOMEDICAL ELECTRODE, P99
[9]  
GLEESON DT, 1996, THESIS TRINITY COLLE
[10]   NON-POLARIZABLE DRY ELECTRODE BASED ON NASICON CERAMIC [J].
GONDRAN, C ;
SIEBERT, E ;
FABRY, P ;
NOVAKOV, E ;
GUMERY, PY .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1995, 33 (03) :452-457