Ankle arthroplasties generate wear particles similiar to knee arthroplasties

被引:31
作者
Kobayashi, A
Minoda, Y
Kadoya, Y
Ohashi, H
Takaoka, K
Saltzman, CL
机构
[1] Osaka City Univ, Sch Med, Dept Orthopaed Surg, Abeno Ku, Osaka 5458585, Japan
[2] Univ Iowa, Dept Biomed Engn, Iowa City, IA 52242 USA
[3] Univ Iowa, Dept Orthopaed Surg, Iowa City, IA 52242 USA
关键词
D O I
10.1097/01.blo.0000132182.54217.da
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Second-generation total ankle arthroplasties have encouraging medium-term results, but the wear of the joint materials is of concern. The aim of the current study was to examine and compare the size, shape, and concentration of polyethylene particles in synovial fluid with total ankle arthroplastics and established posterior-stabilized total knee arthroplasties. Synovial fluid was obtained from 15 patients with well-functioning total ankle arthroplasties and 11 patients with posterior-stabilized total knee arthroplasties at least 6 months after surgery. Polyethylene particles were isolated and analyzed using scanning electron microscopy. Particle size (equivalent circle diameter) in ankles was 0.81 +/- 0.09 mum (mean standard error) and in knees was 0.78 +/- 0.08 mum. Particle shape (aspect ratio) in ankles was 1.57 +/- 0.04 and in knees was 2.30 +/- 0.22. The particle concentration was 1.02 +/- 0.43 x 10(7)/mL in ankles and 1.13 +/- 0.56 x 10(7) /mL in knees, and the particle concentration and size in total ankle arthroplasties were similar to those in total knee arthroplasties. Total ankle arthroplasties generated significantly rounder particles than total knee arthroplasties. These data suggest that the long-term result of total ankle arthroplasty should be as good as posterior-stabilized total knee arthroplasties in terms of polyethylene wear and the prevalence of osteolysis.
引用
收藏
页码:69 / 72
页数:4
相关论文
共 11 条
[1]
TOTAL ANKLE ARTHROPLASTY - A LONG-TERM REVIEW OF THE LONDON HOSPITAL EXPERIENCE [J].
BOLTONMAGGS, BG ;
SUDLOW, RA ;
FREEMAN, MAR .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1985, 67 (05) :785-790
[2]
Easley Mark E, 2002, J Am Acad Orthop Surg, V10, P157
[3]
Font-Rodriguez DE, 1997, CLIN ORTHOP RELAT R, P79
[4]
Hirakawa K, 1996, CLIN ORTHOP RELAT R, P151
[5]
Number of polyethylene particles and osteolysis in total joint replacements - A quantitative study using a tissue-digestion method [J].
Kobayashi, A ;
Freeman, MAR ;
Bonfield, W ;
Kadoya, Y ;
Yamac, T ;
AlSaffar, N ;
Scott, G ;
Revell, PA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (05) :844-848
[6]
Comparison of the response of primary human peripheral blood mononuclear phagocytes from different donors to challenge with model polyethylene particles of known size and dose [J].
Matthews, JB ;
Green, TR ;
Stone, MH ;
Wroblewski, BM ;
Fisher, J ;
Ingham, E .
BIOMATERIALS, 2000, 21 (20) :2033-2044
[7]
Polyethylene wear particles in synovial fluid after total knee arthroplasty [J].
Minoda, Y ;
Kobayashi, A ;
Iwaki, H ;
Miyaguchi, M ;
Kadoya, Y ;
Ohashi, H ;
Yamano, Y ;
Takaoka, K .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (410) :165-172
[8]
Total ankle arthroplasty: a unique design - Two to twelve-year follow-up [J].
Pyevich, MT ;
Saltzman, CL ;
Callaghan, JJ ;
Alvine, FG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1998, 80A (10) :1410-1420
[9]
Saltzman C L, 1999, Instr Course Lect, V48, P263
[10]
STERN SH, 2001, SURG KNEE, V2, P1660