Comparison of harmonic and conventional power Doppler ultrasonography for assessment of slow flow in hyperechoic tissue - Experimental study using a Doppler phantom

被引:9
作者
Kim, AY
Kim, TK
Kim, YH
Han, JK
Choi, BI
机构
[1] Seoul Natl Univ, Coll Med, Dept Radiol, Seoul, South Korea
[2] Seoul Natl Univ, Coll Med, Inst Radiat Med, Seoul, South Korea
关键词
power Doppler ultrasonography; experimental study; harmonic power Doppler imaging;
D O I
10.1097/00004424-200002000-00003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
RATIONALE AND OBJECTIVES. Despite the advantages of depicting slow flow in small vessels, conventional power Doppler ultrasound (US) has a basic limitation, specifically that artifactual power Doppler signals mimic blood flow, especially in hyperechoic tissue. The purpose of this study was to compare harmonic power Doppler US with power Doppler US using a Doppler phantom under various parameter settings, focusing on the assessment of slow flow in the hyperechoic tissue. METHODS. While controlling the flow velocity (5 and 10 cm/s), pulse repetition frequency (500, 700, and 1000 Hz), wall filter (low and medium), and Doppler gain (90%, 96%, and 100%), the authors performed both harmonic Doppler US and power Doppler US by using a Doppler phantom/flow control system, We measured and compared the relative intensities of the Doppler signals (0-250 scale) in both the vessels and hyperechoic tissue-mimicking materials with the two different imaging modalities. RESULTS. Power Doppler US with any combination of the four parameters evaluated depict-ed strong flow signals (mean, 213) that were superior to harmonic Doppler US (mean, 61), Relatively strong artifactual signals within the hyperechoic tissue-mimicking materials were noted on all power Doppler US studies (mean, 106) but nearly none on harmonic Doppler US (mean, 3), The contrast-to-noise ratio of harmonic Doppler US was significantly greater than that of power Doppler US. CONCLUSIONS. Harmonic Doppler US is more useful in assessing slow flow in hyperechoic tissue than power Doppler US because it produces fewer artifactual Doppler signals originating from stationary hyperechoic tissues, which can be misjudged as true signals on power Doppler US.
引用
收藏
页码:105 / 110
页数:6
相关论文
共 12 条
[1]   Power Doppler sonography [J].
Bude, RO ;
Rubin, JM .
RADIOLOGY, 1996, 200 (01) :21-23
[2]   POWER VERSUS CONVENTIONAL COLOR DOPPLER SONOGRAPHY - COMPARISON IN THE DEPICTION OF NORMAL INTRARENAL VASCULATURE [J].
BUDE, RO ;
RUBIN, JM ;
ADLER, RS .
RADIOLOGY, 1994, 192 (03) :777-780
[3]  
BUDE RO, 1996, RADIOLOGY, V26, P109
[4]  
BUNS PN, 1996, ANGIOLOGY, V47, pS63
[5]   Power Doppler US in synovitis: Reality or artifact? [J].
Cardinal, E ;
Lafortune, M ;
Burns, P .
RADIOLOGY, 1996, 200 (03) :868-869
[6]   Power versus conventional color Doppler sonography: Comparison in the depiction of vasculature in liver tumors [J].
Choi, BI ;
Kim, TK ;
Han, JK ;
Chung, JW ;
Park, JH ;
Han, MC .
RADIOLOGY, 1996, 200 (01) :55-58
[7]   INFLUENCE OF VARIOUS INSTRUMENT SETTINGS ON THE FLOW INFORMATION DERIVED FROM THE POWER MODE [J].
JAIN, SP ;
FAN, PH ;
PHILPOT, EF ;
NANDA, NC ;
AGGARWAL, KK ;
MOOS, S ;
YOGANATHAN, AP .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1991, 17 (01) :49-54
[8]  
Nilsson A, 1997, J ULTRAS MED, V16, P679
[9]   Intratumoral vascularity of experimentally induced VX2 carcinoma - Comparison of color doppler sonography, power doppler sonography, and microangiography [J].
Park, KS ;
Choi, BI ;
Won, HJ ;
Seo, JB ;
Kim, SH ;
Kim, TK ;
Han, JK ;
Yeon, KM .
INVESTIGATIVE RADIOLOGY, 1998, 33 (01) :39-44
[10]   COMPARISON OF COLOR DOPPLER ENERGY SONOGRAPHY WITH CONVENTIONAL COLOR DOPPLER SONOGRAPHY IN DETECTION OF FLOW SIGNAL IN PERIPHERAL RENAL-TRANSPLANT VESSELS [J].
PREIDLER, KW ;
SZOLAR, DM ;
UGGOWITZER, M ;
STISKAL, M ;
HORINA, J .
BRITISH JOURNAL OF RADIOLOGY, 1995, 68 (814) :1103-1105