Non-contact respiratory monitoring system using a ceiling-attached microwave antenna

被引:80
作者
Uenoyama, Maki [1 ]
Matsui, Takemi [1 ]
Yamada, Kouske [1 ]
Suzuki, Satoshi [1 ]
Takase, Bonpei [1 ]
Suzuki, Shinya [1 ]
Ishihara, Masayuki [1 ]
Kawakami, Mitsuyuki [1 ]
机构
[1] Tokyo Metropolitan Univ, Dept Management Syst Engn, Hino, Tokyo 1910065, Japan
关键词
respiration; monitoring; non-contact; microwave; nursing care;
D O I
10.1007/s11517-006-0091-8
中图分类号
TP39 [计算机的应用];
学科分类号
081203 [计算机应用技术]; 0835 [软件工程];
摘要
Using a microwave antenna attached to the room ceiling, we conducted non-contact monitoring of respiratory chest wall motions of subjects in bed and covered by a soft comfortable bedding, to measure the vital signs of patients under nursing care in a welfare institution. Long-term vital sign monitoring using electrodes places a heavy burden on monitored individuals. Our non-contact respiratory monitoring system comprises a 1,215 MHz-microwave radar (LDR-1), antenna box attached to the ceiling, and personal computer with analyzing software. The system was tested on eight healthy volunteers (mean age, 25 years; range, 21-44 years) and eight elderly volunteers with some disorders (mean age, 69 years; range, 66-75 years). Respiratory rates of subjects measured using this system correlated with rates measured using respiration sensors (r = 0.97, P < 0.001 for healthy volunteers, r = 0.98, P < 0.0001 for elderly volunteers). The system could monitor subtle changes in respiratory rate, and monitoring respiratory rate increases caused by disorders such as pneumonia will be possible.
引用
收藏
页码:835 / 840
页数:6
相关论文
共 11 条
[1]
AN X-BAND MICROWAVE LIFE-DETECTION SYSTEM [J].
CHEN, KM ;
MISRA, D ;
WANG, H ;
CHUANG, HR ;
POSTOW, E .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (07) :697-701
[2]
Metabolism of the hamster pancreatic carcinogen methyl-2-oxopropylnitrosamme by hamster liver and pancreas [J].
Chen, SC ;
Wang, XJ ;
Zhou, L ;
Kolar, C ;
Lawson, TA ;
Mirvish, SS .
INTERNATIONAL JOURNAL OF PANCREATOLOGY, 2000, 27 (02) :105-112
[3]
AUTOMATIC CLUTTER-CANCELER FOR MICROWAVE LIFE-DETECTION SYSTEMS [J].
CHUANG, HR ;
CHEN, YF ;
CHEN, KM .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1991, 40 (04) :747-750
[4]
Biomedical engineering's contribution to defending the homeland [J].
Kikuchi, M ;
Ishihara, M ;
Matsui, T ;
Wakisaka, H ;
Ashida, H ;
Sato, S ;
Ishizuka, T .
IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 2004, 23 (01) :175-186
[5]
Wandering: A significant problem among community-residing individuals with Alzheimer's disease [J].
Logsdon, RG ;
Teri, L ;
McCurry, SM ;
Gibbons, LE ;
Kukull, WA ;
Larson, EB .
JOURNALS OF GERONTOLOGY SERIES B-PSYCHOLOGICAL SCIENCES AND SOCIAL SCIENCES, 1998, 53 (05) :P294-P299
[6]
Non-contact determination of vital sign alterations in hypovolaemic states induced by massive haemorrhage: an experimental attempt to monitor the condition of injured persons behind barriers or under disaster rubble [J].
Matsui, T ;
Ishizuka, T ;
Takase, B ;
Ishihara, M ;
Kikuchi, M .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2004, 42 (06) :807-811
[7]
A novel apparatus for non-contact measurement of heart rate variability: a system to prevent secondary exposure of medical personnel to toxic materials under biochemical hazard conditions, in monitoring sepsis or in predicting multiple organ dysfunction syndrome [J].
Matsui, T ;
Arai, I ;
Gotoh, S ;
Hattori, H ;
Takase, B ;
Kikuchi, M ;
Ishihara, M .
BIOMEDICINE & PHARMACOTHERAPY, 2005, 59 :S188-S191
[8]
A novel method to prevent secondary exposure of medical and rescue personnel to toxic materials under biochemical hazard conditions using microwave radar and infrared thermography [J].
Matsui, T ;
Hagisawa, K ;
Ishizuka, T ;
Takase, B ;
Ishihara, M ;
Kikuchi, M .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2004, 51 (12) :2184-2188
[9]
MATSUI T, 2006, IN PRESS J MED ENG T
[10]
CITIZEN RESPONSE TO CARDIOPULMONARY EMERGENCIES [J].
MONTGOMERY, WH ;
BROWN, DD ;
HAZINSKI, MF ;
CLAWSEN, J ;
NEWELL, LD ;
FLINT, L .
ANNALS OF EMERGENCY MEDICINE, 1993, 22 (02) :428-434