Selective source blocking for Gamma Knife radiosurgery of trigeminal neuralgia based on analytical dose modelling

被引:11
作者
Li, KL [1 ]
Ma, LJ [1 ]
机构
[1] Univ Maryland, Sch Med, Dept Radiat Oncol, Baltimore, MD 21210 USA
关键词
D O I
10.1088/0031-9155/49/15/010
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
We have developed an automatic critical region shielding (ACRS) algorithm for Gamma Knife radiosurgery of trigeminal neuralgia. The algorithm selectively blocks 201 Gamma Knife sources to minimize the dose to the brainstem while irradiating the root entry area of the trigeminal nerve with 70-90 Gy. An independent dose model was developed to implement the algorithm. The accuracy of the dose model was tested and validated via comparison with the Leksell GammaPlan (LGP) calculations. Agreements of 3% or 3 mm in isodose distributions were found for both single-shot and multiple-shot treatment plans. After the optimized blocking patterns are obtained via the independent dose model, they are imported into the LGP for final dose calculations and treatment planning analyses. We found that the use of a moderate number of source plugs (30-50 plugs) significantly lowered (similar to40%) the dose to the brainstem for trigeminal neuralgia treatments. Considering the small effort involved in using these plugs, we recommend source blocking for all trigeminal neuralgia treatments with Gamma Knife radiosurgery.
引用
收藏
页码:3455 / 3463
页数:9
相关论文
共 11 条
[1]
A spherical dose model for radiosurgery plan optimization [J].
Cho, PS ;
Kuterdem, HG ;
Marks, RJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (10) :3145-3148
[2]
TREATMENT VOLUME SHAPING WITH SELECTIVE BEAM BLOCKING USING THE LEKSELL GAMMA-UNIT [J].
FLICKINGER, JC ;
MAITZ, A ;
KALEND, A ;
LUNSFORD, LD ;
WU, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 19 (03) :783-789
[3]
Knuth DE., 1997, ART COMPUTER PROGRAM
[4]
Stereotactic radiosurgery for the treatment of trigeminal neuralgia [J].
Kondziolka, D ;
Lunsford, LD ;
Flickinger, JC .
CLINICAL JOURNAL OF PAIN, 2002, 18 (01) :42-47
[5]
LEKSELL L, 1971, ACTA CHIR SCAND, V137, P311
[6]
GammaPlan®-Leksell Gamma Knife® radiosurgery treatment planning verification method [J].
Marcu, SM ;
Wu, QJ ;
Pillai, K ;
Weinhous, MS .
MEDICAL PHYSICS, 2000, 27 (09) :2146-2149
[7]
An empirical model for independent dose verification of the Gamma Knife treatment planning [J].
Phaisangittisakul, N ;
Ma, LJ .
MEDICAL PHYSICS, 2002, 29 (09) :1991-1997
[8]
WU A, 1990, INT J RAD ONCOL BIOL, V18
[9]
Morphology-guided radiosurgery treatment planning and optimization for multiple isocenters [J].
Wu, QJ ;
Bourland, JD .
MEDICAL PHYSICS, 1999, 26 (10) :2151-2160
[10]
ZHANG P, 2000, APPL CLIN MED PHYS, V1, P158